Philips And Neusoft Medical Systems Co., Ltd.: The device is the NeuViz Dual series Computed Tomography (CT) Scanner System. The device is identified as NeuViz Dual, PN: 989605651321. Labeling on the device reads in part: "***Neusoft NeuViz Dual MULTI-SLICE CT SCANNER SYSTEM***MANUFACTURED BY PHILIPS AND NEUSOFT MEDICAL SYSTEMS CO., LTD. ADDRESS: NEUSOFT PARK, HUN NAN INDUSTRIAL AREA, SHENYANG 110179, CHINA***". Device is intended to produce cross-section images of head and body by computer reconstruction of x-ray transmission data taken at different angles.

Recall number
Z-0846-2011
Recalling firm
Philips And Neusoft Medical Systems Co., Ltd.
Product
The device is the NeuViz Dual series Computed Tomography (CT) Scanner System. The device is identified as NeuViz Dual, PN: 989605651321. Labeling on the device reads in part: "***Neusoft NeuViz Dual MULTI-SLICE CT SCANNER SYSTEM***MANUFACTURED BY PHILIPS AND NEUSOFT MEDICAL SYSTEMS CO., LTD. ADDRESS: NEUSOFT PARK, HUN NAN INDUSTRIAL AREA, SHENYANG 110179, CHINA***". Device is intended to produce cross-section images of head and body by computer reconstruction of x-ray transmission data taken at different angles.
Status
Terminated
Initiated
2010-02-10
Posted
2011-01-04
Terminated
2011-04-15
Reason
Potential for the CT patient table top to become detached from the carrier due to missing washers.
Root cause
Other
Action
Neusoft Medical Systems Co., China sent an URGENT DEVICE CORRECTION letter dated Feb. 10, 2010 to their distributor in NC. On 02/12/10, The distributor sent the same letter to their consignees. The firm informed this distributor that there was report of the four screws that secure the patient table top to the carrier that did not have the proper washers in place which caused the table top to detach. If the issue would reoccur, there is a potential for the table top to fall and cause patient injury. Customers were instructed to check the patient table top fixing. If it is found that the the assembly is improper customers were instructed to top using the product and contact local Service personnel for repair. Any customer can contact Service Support Department, Neusoft Medical Systems Co, LTD at nms-service@neusoft.com or Service Support Department at helpdesk@neusoft.com for questions about this recall.
Quantity
26 devices
Distribution
Nationwide Distribution including NC
Lot, serial or model codes
Part number: NeuViz Dual 989605651321.  SERIAL NUMBERS:  400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH025EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002, NDHR080003, NDHR080004, NDHR080007, NDHR080012, NDHR080021, NDHR080031, NDHR090003, and NDHR090012.
510(k) numbers
["K062451","K071308"]
PMA numbers
not on file
Product code
JAK
Firm FEI
3003575350
Firm city
Shenyang
Firm state
Liaoning
Firm country
China

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register