Philips Medical Systems (Cleveland) Inc: Computed Tomography X-Ray Systems (Brilliance 64, Model Number: 728231; Brilliance Big Bore, Model Number: 728243; Brilliance iCT, Model Number: 728306; Brilliance iCT SP, Model Number: 728311; and Ingenuity CT, Model Number: 728326); Mfr. by Philips Medical Systems; Cleveland, OH. Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
- Recall number
- Z-0845-2012
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Product
- Computed Tomography X-Ray Systems (Brilliance 64, Model Number: 728231; Brilliance Big Bore, Model Number: 728243; Brilliance iCT, Model Number: 728306; Brilliance iCT SP, Model Number: 728311; and Ingenuity CT, Model Number: 728326); Mfr. by Philips Medical Systems; Cleveland, OH. Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
- Status
- Terminated
- Initiated
- 2011-10-07
- Posted
- 2012-01-24
- Terminated
- 2013-06-18
- Reason
- The Common Image Reconstruction System (CIRS) on the affected Brilliance Systems encounters an error and requires reboot of the system. The unit failed to initialize axial scan due to timeout that occurred waiting for "READY" state
- Root cause
- Software design
- Action
- On 10/10/2011 the firm sent Urgent - Medical Device Correction letters to their customers. The letters identified the affected products, the problem, the hazard involved, the action to be taken by the customer, and actions planned by Philips. Customers were to reboot their systems if the two possible scenarios occurred. Philips Healthcare plans on installing the update through a Field Change Order free of charge. For further information or support concerning the issue, contact a local Philips representative or local Philips Healthcare office. In North America and Canada, contact Customer Care Solutions Center at 1-800-722-9377, option 5.
- Quantity
- 548 units
- Distribution
- Worldwide Distribution -- USA, including the states of: AL, AZ, CA, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NM, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, VT, WA & WV and the countries of: ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BANGLADESH, BELGIUM, Brazil, CHINA, COLUMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FINLAND, FRANCE, GEORGIA, GERMANY, HUNGARY, ICELAND, INDIA, INDONESIA, IRAQ, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KOREA, LUXEMBOURG, MALAYSIA, MALI, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PHILIPPINES, POLAND, QATAR, RUSSIA, SINGAPORE, SLOVENIA, SOUTH KOREA, SPAIN, SUDAN, SWEDEN, SWITZERLAND, SYRIAN ARAB REPUBLIC, TAIWAN, THAILAND, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, UZBEKISTAN & VIETNAM.
- Lot, serial or model codes
- Model #'s - Brilliance 64: 728231, Ingenuity CT: 728326, iCT: 728306; iCT SP: 728311 & Brilliance Big Bore: 728243
- 510(k) numbers
- ["K033326","K033357","K060937"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 1525965
- Firm city
- Cleveland
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28