Skip to the content

FDA Medical Device Recalls (openFDA)

Arrow International Inc: Arrow Arterial Catherization Kit-20 Ga. x 4.45 cm Radiopaque Polyurethane over 22 Ga. TW Introducer Needle with Integral .018" (0.46 mm) dia. Spring-Wire Guide Product Code: ASK-04020-PMC The Arrow Arterial Catheterization device permits access to the peripheral arterial circulation or to other small vessels.

Recall number
Z-0844-2020
Recalling firm
Arrow International Inc
Product
Arrow Arterial Catherization Kit-20 Ga. x 4.45 cm Radiopaque Polyurethane over 22 Ga. TW Introducer Needle with Integral .018" (0.46 mm) dia. Spring-Wire Guide Product Code: ASK-04020-PMC The Arrow Arterial Catheterization device permits access to the peripheral arterial circulation or to other small vessels.
Status
Open, Classified
Initiated
2019-12-12
Posted
not on file
Terminated
not on file
Reason
The lidstock label incorrectly displays a MR (Magnetic Resonance ) Conditional symbol (instead of a MR Unsafe symbol) in reference to a component of the kit
Root cause
Labeling Change Control
Action
Arrow International issued an " Urgent Medical Device Recall Notice" to its US consignee on 12th Dec 2019 via FedEx 2-day mail. The letter states reason for recall, health risk and action to take: If you have affected stock in inventory, immediately discontinue use and quarantine the products. 2. Please complete the enclosed Acknowledgement Form as soon as possible indicating whether you do or do not have stock and fax it to 1-855-419-8507 or email to recalls@teleflex.com. A customer service representative will contact you with a Return Goods Authorization (RGA) number and provide instructions for the return of products. If you have no affected stock, please complete the enclosed Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Questions contact your local sales representative or Customer Service at 1-866-396-2111.
Quantity
80 eaches
Distribution
US distribution to SC
Lot, serial or model codes
Lot/Batch Number: PMC 23F19H0224
510(k) numbers
["K810675"]
PMA numbers
not on file
Product code
DQX
Firm FEI
3015859709
Firm city
Reading
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register