FDA Medical Device Recalls (openFDA)

Ortho Clinical Diagnostics Inc: Vitros Myoglobin Calibrators, Product Code 6801043 For use in the calibration of the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System for the quantitative measurement of myoglobin in human serum and plasma

Recall number
Z-0844-2019
Recalling firm
Ortho Clinical Diagnostics Inc
Product
Vitros Myoglobin Calibrators, Product Code 6801043 For use in the calibration of the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System for the quantitative measurement of myoglobin in human serum and plasma
Status
Terminated
Initiated
2018-11-29
Posted
not on file
Terminated
2024-02-06
Reason
A stability issue in the calibrators may cause calibration failures or atypical Quality Control results following a successful calibration.
Root cause
Under Investigation by firm
Action
On November 29, 2018, the firm sent a customer letter all affected customers. All consignees were notified of the issue and informed of revised instructions for VITROS Myoglobin Calibrators: - Reconstitute the calibrators using 1.0 mL distilled water, as per current instructions. - Allow reconstituted calibrator vials to stand at room temperature for 2 hours and then proceed with calibration within 2 hours (4 hours total) - Discard reconstituted calibrators after 4 hours. - Acknowledge the information contained in this notification by completing and submitting the Confirmation of Receipt form no later than December 14, 2018. - Post this notification by each VITROS System at your facility that processes this assay or with the user documentation. - Please forward this notification if the product was distributed outside of your facility. If you have questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.
Quantity
4030
Distribution
Distributed nationwide domestically. Foreign distribution to Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Singapore, UK, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, and the Netherlands.
Lot, serial or model codes
Lots 1271 through 1310 (expired) Lots 1315 & Above (in-date)
510(k) numbers
["K033300"]
PMA numbers
not on file
Product code
JIS
Firm FEI
2250051
Firm city
Raritan
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register