FDA Medical Device Recalls (openFDA)
GE Healthcare: GE Healthcare Proteus XR/a Radiographic X-ray System, Model Number 2259976; previously marketed as the Silhouette FC, GE Healthcare, Waukesha, WI 53188.
- Recall number
- Z-0842-2008
- Recalling firm
- GE Healthcare
- Product
- GE Healthcare Proteus XR/a Radiographic X-ray System, Model Number 2259976; previously marketed as the Silhouette FC, GE Healthcare, Waukesha, WI 53188.
- Status
- Terminated
- Initiated
- 2007-11-01
- Posted
- 2008-06-10
- Terminated
- 2011-12-18
- Reason
- The warning label that is required by 21 CFR 1020.30(j) was not on the control console of the Proteus XR/a Radiographic X-ray System, Model 2259976.
- Root cause
- Process control
- Action
- GE issued a Field Modification Instruction to their field service representatives, released in December 2007, with instructions to replace the labels on the console with the appropriate labeling. Consignees will be visited by GE field service representatives beginning December 2007. All consignees will have the required labels affixed to their units by July 31, 2008.
- Quantity
- 400 Units
- Distribution
- Worldwide Distribution, including USA, Canada, Mexico, France, Hong Kong, Italy, Korea, China, Poland, UK, New Zealand, Saudi Arabia, Taiwan, Turkey, and Australia.
- Lot, serial or model codes
- 00000010838HL1, 00000011061HL9, 00000011067HL6, 00000011185HL6, 00000012500HL5, 00000012667HL2, 00000013129HL2, 00000013372HL8, 00000013385HL0, 00000013543HL4, 00000013557HL4, 00000013800HL8, 00000014099HL6, 00000014104HL4, 00000014319HL8, 00000014327HL1, 00000014667HL0, 00000014677HL9, 00000015076HL3, 00000015383HL3, 00000015386HL6, 00000015400HL5, 00000015738HL8, 00000015744HL6, 00000015745HL3, 00000015750HL3, 00000015760HL2, 00000016174HL5, 00000016383HL2, 00000016386HL5, 00000016696HL7, 00000016707HL2, 00000016708HL0, 00000016709HL8, 00000016930HL0, 00000016954HL0, 00000016955HL7, 00000017119HL9, 00000017130HL6, 00000017131HL4, 00000017476HL3, 00000017477HL1, 00000017478HL9, 00000017480HL5, 00000017490HL4, 00000017492HL0, 00000017493HL8, 00000017599HL2, 00000017604HL0, 00000017608HL1, 00000017610HL7, 00000017613HL1, 00000017616HL4, 00000017788HL1, 00000017801HL2, 00000018476HL2, 00000018479HL6, 00000018483HL8, 00000019041HL3, 00000019048HL8, 00000019095HL9, 00000019098HL3, 00000019104HL9, 00000019110HL6, 00000019241HL9, 00000019244HL3, 00000019413HL4, 00000019419HL1, 00000019489HL4, 00000019496HL9, 00000019499HL3, 00000019501HL6, 00000019502HL4, 00000019669HL1, 00000019801HL0, 00000019886HL1, 00000019894HL5, 00000019959HL6, 00000020091HL5, 00000020092HL3, 00000020239HL0, 00000020243HL2, 00000020374HL5, 00000020967HL6, 00000021064HL1, 00000021197HL9, 00000021201HL9, 00000021204HL3, 00000021343HL9, 00000021344HL7, 00000021349HL6, 00000021552HL5, 00000021911HL3, 00000021921HL2, 00000021938HL6, 00000021953HL5, 00000022033HL5, 00000022333HL9, 00000022337HL0, 00000022412HL1, 00000022477HL4, 00000022600HL1, 00000022795HL9, 00000022931HL0, 00000023116HL7, 00000023122HL5, 00000023126HL6, 00000023412HL0, 00000023415HL3, 00000024386HL5, 00000024483HL0, 00000024588HL6, 00000024939HL1, 00000024940HL9, 00000025426HL8, 00000025475HL5, 00000025476HL3, 00000025477HL1, 00000025572HL9, 00000027104HL9, 00000027867HL1, 00000028057HL8, 00000028453HL9, 00000028524HL7, 00000028845HL6, 00000029098HL1, 00000029145HL0, 00000029502HL2, 00000029550HL1, 00000029634HL3, 00000029640HL0, 00000029969HL3, 00000030108HL5, 00000030113HL5, 00000030229HL9, 00000030232HL3, 00000030236HL4, 00000030633HL2, 00000030848HL6, 00000030920HL3, 00000030921HL1, 00000030976HL5, 00000031172HL0, 00000031175HL3, 00000031565HL5, 00000031702HL4, 00000031811HL3, 00000031928HL5, 00000031929HL3, 00000031930HL1, 00000032309HL7, 00000032311HL3, 00000032317HL0, 00000032319HL6, 00000032322HL0, 00000032850HL0, 00000032912HL8, 00000032916HL9, 00000032919HL3, 00000032920HL1, 00000033330HL2, 00000033388HL0, 00000033391HL4, 00000033408HL6, 00000033437HL5, 00000033509HL1, 00000033679HL2, 00000033681HL8, 00000033682HL6, 00000033727HL9, 00000033731HL1, 00000033884HL8, 00000033970HL5, 00000033971HL3, 00000033983HL8, 00000033984HL6, 00000034218HL8, 00000034420HL0, 00000034469HL7, 00000034470HL5, 00000034520HL7, 00000034644HL5, 00000034694HL0, 00000034762HL5, 00000034764HL1, 00000034913HL4, 00000034938HL1, 00000035052HL0, 00000035101HL5, 00000035159HL3, 00000035191HL6, 00000035244HL3, 00000035246HL8, 00000035247HL6, 00000035329HL2, 00000035445HL6, 00000035544HL6, 00000035547HL9, 00000035632HL9, 00000035641HL0, 00000036038HL8, 00000036068HL5, 00000036071HL9, 00000036100HL6, 00000036126HL1, 00000036128HL7, 00000036265HL7, 00000036362HL2, 00000036363HL0, 00000036431HL5, 00000036624HL5, 00000036625HL2, 00000036674HL0, 00000036698HL9, 00000036702HL9, 00000036882HL9, 00000036983HL5, 00000037104HL7, 00000037123HL7, 00000037351HL4, 00000037352HL2, 00000037353HL0, 00000037396HL9, 00000037596HL4, 00000037919HL8, 00000038423HL0, 00000038771HL2, 00000038772HL0, 00000039125HL0, 00000039581HL4, 00000039612HL7, 00000040366HL7, 00000040493HL9, 00000040728HL8, 00000040865HL8, 00000041290HL8, 00000041291HL6, 00000041511HL7, 00000041768HL3, 00000042367HL3, 00000042649HL4, 00000042650HL2, 00000042651HL0, 00000042701HL3, 00000042880HL5, 00000043029HL8, 00000043184HL1, 00000043265HL8, 00000043665HL9, 00000043727HL7.00000043747HL5, 00000043751HL7, 00000044548HL6, 00000045475HL1, 00000045651HL7, 00000015925HL1, 00000019969HL5, 00000024476HL4, 00000027957HL0, 00000030025HL1, 00000030028HL5, 00000032711HL4, 00000035548HL7, 00000036428HL1, 00000036883HL7, 00000020371HL1, 00000021582HL2, 00000031922HL8, 00000033316HL1, 00000042706HL2, 00000020735HL7, 00000022513HL6, 00000022926HL0, 00000024376HL6, 00000031909HL5, 00000040521HL7, 00000041017HL5, 00000011544HL4, 00000014336HL2, 00000016171HL1, 00000016545HL6, 00000017123HL1, 00000017483HL9, 00000023189HL4, 00000028833HL2, 00000030875HL9, 00000030924HL5, 00000031916HL0, 00000031918HL6, 00000032909HL4, 00000033730HL3, 00000033974HL7, 00000034268HL3, 00000036165HL9, 00000042001HL8, 00000016177HL8, 00000031554HL9, 00000036626HL0, 00000013383HL5, 00000014323HL0, 00000019485HL2, 00000020232HL5, 00000010836HL5, 00000019679HL0, 00000023546HL5, 00000023547HL3, 00000030926HL0, 00000041996HL0, 00000042388HL9, 00000042537HL1, 00000011063HL5, 00000015391HL6, 00000017138HL9, 00000021058HL3, 00000030918HL7, 00000040086HL1, 00000010328HL3, 00000013375HL1, 00000014335HL4, 00000015065HL6, 00000015081HL3, 00000015747HL9, 00000016371HL7, 00000016701HL5, 00000017125HL6, 00000017481HL3, 00000019482HL9, 00000019486HL0, 00000019678HL2, 00000019680HL8, 00000019788HL9, 00000020081HL6, 00000020224HL2, 00000021639HL0, 00000022032HL7, 00000023118HL3, 00000023181HL1, 00000024432HL7, 00000029259HL9, 00000031908HL7, 00000031914HL5, 00000032231HL3, 00000032897HL1, 00000033724HL6, 00000033725HL3, 00000033968HL9, 00000034260HL0, 00000034459HL8, 00000034462HL2, 00000034783HL1, 00000035377HL1, 00000035378HL9, 00000036563HL5, 00000037254HL0, 00000038190HL5, 00000016175HL2, 00000019042HL1, 00000013796HL8, 00000011182HL3, 00000032313HL9, 00000012475HL0, 00000020222HL6, 00000029743HL2, 00000033726HL1, 00000020732HL4, 00000020737HL3, 00000012674HL8, 00000021551HL7, 00000031689HL3, 00000010651HL8, 00000015757HL8, 00000022520HL1, 00000015082HL1, 00000015153HL0, 00000021644HL0, 00000013373HL6, 00000015158HL9, 00000022606HL8, 00000022607HL6, 00000024489HL2, 00000024590HL2, 00000024671HL0, 00000024689HL2, 00000030446HL9, 00000030460HL0, 00000030927HL8, 00000036429HL9, 00000015387HL4, 00000039893HL3, 00000039895HL8, 00000039962HL6, 00000040002HL8, 00000040072HL1, 00000040164HL6, 00000040167HL9, 00000022402HL2, 00000036266HL5, 00000036914HL0, 00000037252HL4, 00000013564HL0.
- 510(k) numbers
- ["K993090"]
- PMA numbers
- not on file
- Product code
- KPR
- Firm FEI
- 2126677
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28