Arrow International Inc: The Arrow Trerotola Percutaneous Thrombolytic Device (Product Codes: PT-65709-WC, PT-65709-W, PT-65709-HFWC, and PT-12709-WC) consists of a catheter with an expandable fragmentation basket attached to a through lumen torque cable and an outer catheter sheath. The catheter is used in conjunction with a low-speed, battery operated rotator unit to treat thrombosed hemodialysis grafts and AV fistulas. The macerating mechanism is done by 9 mm basket (made up of four stranded stainless steel wires) that rotates at 3000 rpm when rotator is activated. The current device is 7 Fr in size and comes in two lengths -- 65 cm and 120 cm. This device is used by inserting the catheter into the occluded graft or fistula via introducer sheath and then activating the rotator unit.
- Recall number
- Z-0842-05
- Recalling firm
- Arrow International Inc
- Product
- The Arrow Trerotola Percutaneous Thrombolytic Device (Product Codes: PT-65709-WC, PT-65709-W, PT-65709-HFWC, and PT-12709-WC) consists of a catheter with an expandable fragmentation basket attached to a through lumen torque cable and an outer catheter sheath. The catheter is used in conjunction with a low-speed, battery operated rotator unit to treat thrombosed hemodialysis grafts and AV fistulas. The macerating mechanism is done by 9 mm basket (made up of four stranded stainless steel wires) that rotates at 3000 rpm when rotator is activated. The current device is 7 Fr in size and comes in two lengths -- 65 cm and 120 cm. This device is used by inserting the catheter into the occluded graft or fistula via introducer sheath and then activating the rotator unit.
- Status
- Terminated
- Initiated
- 2005-03-18
- Posted
- 2005-05-13
- Terminated
- 2006-11-22
- Reason
- The orange Pebax lumen located within the catheter basket may separate from the catheter basket sleeve permitting the lumen to become dislodged during use.
- Root cause
- Other
- Action
- The firm began notification of consignees on 3/23/2005 via certified letter.
- Quantity
- 2418 UNITS
- Distribution
- Product was distributed nationwide to dealers, hospitals, sales representatives, and subsidiaries (distributors). The product was also distributed to Australia, Canada, Colombia, Czech Republic, France, Germany, Israel, Italy, Japan, Netherlands, Russia, South Africa, South Korea, Spain, Sweden, Taiwan, Turkey, United Kingdom. There are no government accounts.
- Lot, serial or model codes
- All units with Lot Numbers beginning with: MF311; MF312; MF401; MF402; MF403; MF404; MF405; MF406; MF407; MF408; MF409; MF410; MF411; MF412; MF501.
- 510(k) numbers
- ["K011056"]
- PMA numbers
- not on file
- Product code
- DXE
- Firm FEI
- 3015859709
- Firm city
- Reading
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28