Skip to the content

FDA Medical Device Recalls (openFDA)

GE OEC Medical Systems, Inc: GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.

Recall number
Z-0840-2015
Recalling firm
GE OEC Medical Systems, Inc
Product
GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.
Status
Terminated
Initiated
2014-09-15
Posted
2015-01-13
Terminated
2015-05-20
Reason
mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).
Root cause
Radiation Control for Health and Safety Act
Action
Planned action(s) to repair defect or to bring product into compliance: GE Healthcare will, without charge, perform a mA calibration on all affected systems in the USA to bring the product back into compliance. The details of which will be included in a subsequent communication to customers or through a GE Healthcare field engineer site visit. Should customers have any questions or concerns regarding these corrections, please do not hesitate to contact the service team for further information at 800-874-7378 option 8.
Quantity
61 units installed in US
Distribution
Worldwide Distribution
Lot, serial or model codes
GE OEC Brivo 715 Prime Serial Numbers:  B2S1301 6 B2S1301 8 B2S1302 1 B2S13022 B2S13025 B2S13039 B2S13060 B2S13063 B2S13065 B2S14001 B2S14007 B2S14022 B2S14039 B2S1404 0 B2S1404 5 B2S1406 5 B2S1406 6
510(k) numbers
["K123603"]
PMA numbers
not on file
Product code
OXO
Firm FEI
1720753
Firm city
Salt Lake City
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28