FDA Medical Device Recalls (openFDA)
Terumo Cardiovascular Systems Corporation: Roller Pump, 6 Inch Diameter The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
- Recall number
- Z-0840-2012
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Roller Pump, 6 Inch Diameter The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
- Status
- Terminated
- Initiated
- 2012-01-13
- Posted
- 2012-01-24
- Terminated
- 2012-12-03
- Reason
- Terumo Cardiovascular Systems (Terumo CVS) discovered and rejected a single digital encoder cable assembly during incoming inspection because it had a loose wire. The assembly is used in the 4" and the 6" roller pumps for the Terumo Advanced Perfusion System 1.
- Root cause
- Nonconforming Material/Component
- Action
- TERUMO sent an Urgent Medical Device Correction letter dated January 13, 2012, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to review the Medical Device Correction notice. Assure that all users are aware of the notice. Confirm receipt of the communication by faxing the attached Customer Response Form to 1-900-292-6551. Roller pumps with affected cables will have the cables assembly replaced. For questions regarding this recall call 1-800-521-2818.
- Quantity
- 44
- Distribution
- Worldwide Distribution - USA including MD, WI, and IA and the countries of Canada, Mexico, and Vietnam
- Lot, serial or model codes
- Catalog Number 816571; Serial Numbers 6621 - 6640
- 510(k) numbers
- ["K022947"]
- PMA numbers
- not on file
- Product code
- DTQ
- Firm FEI
- 1828100
- Firm city
- Ann Arbor
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28