Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc Endoscopy Division: Smith & Nephew, 420XL Xenon Light Source, 180 Watt; for endoscopic use. Catalog/Part #72200510; manufactured for Smith & Nephew Endoscopy, Andover, MA 01810.

Recall number
Z-0840-2007
Recalling firm
Smith & Nephew, Inc Endoscopy Division
Product
Smith & Nephew, 420XL Xenon Light Source, 180 Watt; for endoscopic use. Catalog/Part #72200510; manufactured for Smith & Nephew Endoscopy, Andover, MA 01810.
Status
Terminated
Initiated
2007-04-09
Posted
2007-06-07
Terminated
2008-09-02
Reason
Several units were manufactured using a power supply that contains a 1 amp fuse where a 5 amp fuse is required. When the unit's power supply draws more that 1A, the fuse will fail causing the light to cease operating. The problem can occur anytime: from the initial turning on (powering up), to anytime during use.
Root cause
Nonconforming Material/Component
Action
Notification sent 04/09/07 via FedEx to distribution centers, sales reps and user facilities. Users asked to send light source to Smith & Nephew GMBH Service Center in Germany for repair (contact Ute Dieling at +49 (0) 7462-208-313).
Quantity
67 units.
Distribution
Product sent to distributors in Argentina, Austria, Chile, Denmark, Dubai, France, Greece, Hong Kong, India, Italy, Japan, Korea, Malaysia, Norway, South Africa, Spain, Sweden and United Kingdom. No US distribution.
Lot, serial or model codes
Serial #s: 420L-1263, 420L-1264, 420L-1265, 420L-1266, 420L-1267, 420L-1268, 420L-1269, 420L-1270, 420L-1271, 420L-1272, 420L-1273, 420L-1275, 420L-1277, 420L-1280, 420L-1281, 420L-1282, 420L-1283, 420L-1284, 420L-1285, 420L-1286, 420L-1287, 420L-1288, 420L-1292, 420L-1293, 420L-1294, 420L-1295, 420L-1296, 420L-1298, 420L-1300, 420L-1301, 420L-1303, 420L-1304, 420L-1305, 420L-1306, 420L-1307, 420L-1308, 420L-1309, 420L-1310, 420L-1311, 420L-1312, 420L-1313, 420L-1314, 420L-1315, 420L-1316, 420L-1317, 420L-1318, 420L-1319, 420L-1320, 420L-1321, 420L-1322, 420L-1344, 420L-1345, 420L-1346, 420L-1347, 420L-1348, 420L-1349, 420L-1351, 420L-1352, 420L-1353, 420L-1354, 420L-1355, 420L-1370, 420L-1403, 420L-1404, 420L-1409, 420L-1422 and 420L-1426.
510(k) numbers
["K994084"]
PMA numbers
not on file
Product code
FFS
Firm FEI
1643264
Firm city
Oklahoma City
Firm state
OK
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.