EV3 Neurovascular: Silver Speed Hydrophilic Guidewire, for placement of vascular catheters. Model # 103-0601-200.

Recall number
Z-0839-2007
Recalling firm
EV3 Neurovascular
Product
Silver Speed Hydrophilic Guidewire, for placement of vascular catheters. Model # 103-0601-200.
Status
Terminated
Initiated
2007-04-05
Posted
2007-06-07
Terminated
2012-08-23
Reason
Two customer complaints were received from physicians in Japan indicating the hydrophilic coating was observed flaking from the guidewire while immersed in hydrating media following use during aneurysm embolization procedures.
Root cause
Other
Action
A recall plan was developed for 100% recovery of guidewires within affected lot numbers. The plan includes notification to U.S., European, and Rest-of-World customers and distributors. A recall notification letter was prepared 04/05/2007 which summarizes the coating problem and potential risks, identifies the affected lot numbers and quantities shipped to the customer, and provides instructions for return of unused guidewires.
Quantity
2120
Distribution
Worldwide, including USA, Italy, United Kingdom, Switzerland, Germany, Belgium, South Africa, Canada, Slovenia, China, Austria, France, Spain, Jordan, Mexico, Portugal, Colombia, Greece, Brazil, Venezula, Ireland, Australia, Malaysia, Saudi Arabia, South Korea, Lithuania, Netherlands, Hungary, and United Arab Emirates
Lot, serial or model codes
US: Lot # : 2295725; 2382836; 2355074; 2390635; 2348116; 2355074; 2374619; 2216295; 2257803;2390635; 2342166; ;2390635;2348116; 2310664; 2382836; 2257803; 2216295; 2310664; 2390635; 2247199; 2247200; 2221847; 2247200; 2295725; 2310664; 2221845; 2310664; 2216296; 2257803; 2342166;   European Union & Rest-Of-World Lot #:  2374619 ; 2209708;2216295; 2247200; 2310664; 2348116; 2221845 ; 2355074; 2247200; 2310664; 2216295; 2166877; 2247199; 2209708; 2367901; 2370509; 2374619; 2216295; 2163496; 2174470; 2295725; 2370509; 2376907; 2209708; 2295725; 2310664; 2221845; 2247200; 2216291; 2390635; 2360938; 2370509; 2376907; 2310664; 2355074; 2163496; 2276405; 2370508; 2221847; 2257803; 2390635; 2257803; 2370509; 2200727;
510(k) numbers
["K982543"]
PMA numbers
not on file
Product code
MKI
Firm FEI
3004904811
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register