Leica Microsystems, Inc.: PROVEO 8 Ophthalmic Microscope System optical instrument for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.

Recall number
Z-0838-2019
Recalling firm
Leica Microsystems, Inc.
Product
PROVEO 8 Ophthalmic Microscope System optical instrument for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.
Status
Terminated
Initiated
2017-09-25
Posted
not on file
Terminated
2020-01-21
Reason
The observed issues can lead to unexpected interruptions of the optics carrier focus drive movement, intermittent stops of the X-Y coupling movement or sporadic failures in the proper on-screen display functionality of the HDR recording unit. Additionally, an anomaly in the device software may also lead to the unexpected behavior that when a user releases a handle or foot switch button this may not immediately end the actuated functionality. This behavior can only occur in unlikely workflows where a user changes a user profile or a mode change is initiated at the same time as a functional button is activated.
Root cause
Other
Action
The firm initiated their field notification by email on 09/25/2017. The firm provided details on the issues and requested the user to take the following actions: "If you notice any of the anomalies as described above before being contacted by a Leica Microsystems representative, please contact your Leica representative for relevant repair of your PROVEO 8. You can continue to use the system pending the intervention of the appropriate Leica Microsystems representative. However, please ensure that a functional check is performed and the system is properly balanced prior to surgery as required by and described in chapters 17.1 Checklist before operation and 7.7.1 Balancing the parallelogram of the user manual. During surgery, ensure that no functional switches are activated at the same time as a user profile or mode is changed. During the start-up of the system, ensure that no functional switches are activated until the system boots up."
Quantity
142 units
Distribution
Worldwide
Lot, serial or model codes
Serial Numbers: 190216001; 060516001; 280916001; 061216001; 091216001; 301015001; 270516001; 180716001; 180716002; 190716001; 141016001; 211016001; 241016001; 261016001; 030317001; 060317001; 011216001; 020317001; 130716002; 100316001; 230216001; 041115001; 021115001; 270416001; 130516001; 160516001; 240516001; 220217001; 100417001; 131115001; 071216001; 100816001; 300816001; 231116001; 121216001; 100816001; 130616001; 081216001; 131216001; 140317001; 310816001; 160816001; 130117001; 060217001; 230317001; 280317001; 050417001; 170117001; 190416001; 070415004; 050516001; 090516001; 170516001; 020816001; 040816001; 111016001; 131116001; 150217001; 211116001; 090217001; 081116001; 020616001; 050117001; 031216001; 020416001; 010317001; 280416001; 271216001; 251115001; 250316001; 231216001; 230916001; 230217001; 260416001; 030616001; 080616001; 080316001; 250116001; 181016001; 220316001; 130716002; 201116001; 200217001; 180416001; 201216001; 170117002; 150116001; 121115001; 110316001; 060616001; 270217001; 090317001; 160317001; 190217001; 140716002; 101016001; 161115001; 190116001; 130116001; 010316001; 020217001; 121016001; 141016001; 161216001; 171216001; 180116001; 260516001; 300516001; 200716001; 260716001; 191116001; 171116001; 251116001; 131216001; 220416001; 281016001; 011116001; 270116001; 060416001; 231115001; 301015003; 301015002; 280116001; 150216001; 180216001; 290416001; 040516001; 190516001; 230516001; 250516001; 280616001; 300616001; 200616001; 230616001; 140716001; 130716001; 010916001; 050916001; 100117001; 260116001; 101116001; 041216001
510(k) numbers
not on file
PMA numbers
not on file
Product code
HRM
Firm FEI
3000210133
Firm city
Buffalo Grove
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register