FDA Medical Device Recalls (openFDA)
Leica Microsystems, Inc.: PROVEO 8 Ophthalmic Microscope System optical instrument for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.
- Recall number
- Z-0838-2019
- Recalling firm
- Leica Microsystems, Inc.
- Product
- PROVEO 8 Ophthalmic Microscope System optical instrument for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.
- Status
- Terminated
- Initiated
- 2017-09-25
- Posted
- not on file
- Terminated
- 2020-01-21
- Reason
- The observed issues can lead to unexpected interruptions of the optics carrier focus drive movement, intermittent stops of the X-Y coupling movement or sporadic failures in the proper on-screen display functionality of the HDR recording unit. Additionally, an anomaly in the device software may also lead to the unexpected behavior that when a user releases a handle or foot switch button this may not immediately end the actuated functionality. This behavior can only occur in unlikely workflows where a user changes a user profile or a mode change is initiated at the same time as a functional button is activated.
- Root cause
- Other
- Action
- The firm initiated their field notification by email on 09/25/2017. The firm provided details on the issues and requested the user to take the following actions: "If you notice any of the anomalies as described above before being contacted by a Leica Microsystems representative, please contact your Leica representative for relevant repair of your PROVEO 8. You can continue to use the system pending the intervention of the appropriate Leica Microsystems representative. However, please ensure that a functional check is performed and the system is properly balanced prior to surgery as required by and described in chapters 17.1 Checklist before operation and 7.7.1 Balancing the parallelogram of the user manual. During surgery, ensure that no functional switches are activated at the same time as a user profile or mode is changed. During the start-up of the system, ensure that no functional switches are activated until the system boots up."
- Quantity
- 142 units
- Distribution
- Worldwide
- Lot, serial or model codes
- Serial Numbers: 190216001; 060516001; 280916001; 061216001; 091216001; 301015001; 270516001; 180716001; 180716002; 190716001; 141016001; 211016001; 241016001; 261016001; 030317001; 060317001; 011216001; 020317001; 130716002; 100316001; 230216001; 041115001; 021115001; 270416001; 130516001; 160516001; 240516001; 220217001; 100417001; 131115001; 071216001; 100816001; 300816001; 231116001; 121216001; 100816001; 130616001; 081216001; 131216001; 140317001; 310816001; 160816001; 130117001; 060217001; 230317001; 280317001; 050417001; 170117001; 190416001; 070415004; 050516001; 090516001; 170516001; 020816001; 040816001; 111016001; 131116001; 150217001; 211116001; 090217001; 081116001; 020616001; 050117001; 031216001; 020416001; 010317001; 280416001; 271216001; 251115001; 250316001; 231216001; 230916001; 230217001; 260416001; 030616001; 080616001; 080316001; 250116001; 181016001; 220316001; 130716002; 201116001; 200217001; 180416001; 201216001; 170117002; 150116001; 121115001; 110316001; 060616001; 270217001; 090317001; 160317001; 190217001; 140716002; 101016001; 161115001; 190116001; 130116001; 010316001; 020217001; 121016001; 141016001; 161216001; 171216001; 180116001; 260516001; 300516001; 200716001; 260716001; 191116001; 171116001; 251116001; 131216001; 220416001; 281016001; 011116001; 270116001; 060416001; 231115001; 301015003; 301015002; 280116001; 150216001; 180216001; 290416001; 040516001; 190516001; 230516001; 250516001; 280616001; 300616001; 200616001; 230616001; 140716001; 130716001; 010916001; 050916001; 100117001; 260116001; 101116001; 041216001
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HRM
- Firm FEI
- 3000210133
- Firm city
- Buffalo Grove
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28