Siemens Healthcare Diagnostics, Inc.: Dimension Vista 500 Intelligent Lab System or Dimension Vista 1500 Intelligent Lab System: An in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids.

Recall number
Z-0838-2015
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Dimension Vista 500 Intelligent Lab System or Dimension Vista 1500 Intelligent Lab System: An in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids.
Status
Terminated
Initiated
2014-11-05
Posted
2014-12-20
Terminated
2017-10-06
Reason
Customer complaints of discrepant flagged and/or un-flagged patient and QC results. Reagent probe performance may decline more quickly than anticipated and have an effect on assays with similar formulations when processed sequentially on the same reagent server.
Root cause
Device Design
Action
An Urgent Medical Device Correction, dated November 2014, was issued on November 4th, 2014 to inform customers that Siemens Healthcare Diagnostics has confirmed that reagent probe performance on Dimension Vista systems may decline more quickly than anticipated and have an effect on assays with similar formulations when processed sequentially on the same reagent server. Instructions for replacing an associated reagent probe were provided to resolve the issue. Customers are to retain the letter for their laboratory records and forward it to anyone who has received the affected product. Customers are also to complete and return the Field Correction Effectiveness Check Form. Customer should direct their questions to their Siemens Customer Care Center - Technical Solutions at 1-800-441-9250 or their local Siemens technical support representative.
Quantity
1,949
Distribution
Worldwide Distribution -- Austria, Australia, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Malaysia, Netherlands, New Zealand, Norway, Portugal, Republic Korea, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Switzerland, United Kingdom, and United States.
Lot, serial or model codes
Material Nos.: 10284473, 10488224,  10444801, and 10444802
510(k) numbers
["K051087"]
PMA numbers
not on file
Product code
JJE
Firm FEI
2517506
Firm city
Newark
Firm state
DE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register