Skip to the content

FDA Medical Device Recalls (openFDA)

Meridian Bioscience Inc: Revogene, For In Vitro Diagnostic use in performing nucleic acid testing in clinical laboratories. Meridian branded Revogene: Catalog Number 610210 GenePOC branded Revogene: Catalog Number 130840

Recall number
Z-0836-2022
Recalling firm
Meridian Bioscience Inc
Product
Revogene, For In Vitro Diagnostic use in performing nucleic acid testing in clinical laboratories. Meridian branded Revogene: Catalog Number 610210 GenePOC branded Revogene: Catalog Number 130840
Status
Open, Classified
Initiated
2022-02-15
Posted
not on file
Terminated
not on file
Reason
The photomultiplier tube, may have compromised functionality in some instruments. As a result, the user is likely to observe an increase in false-positive or indeterminate results.
Root cause
Component design/selection
Action
A Letter dated 02/15/2022 sent via FedEx notified users of the risk of false-positives and provided instructions on how to recognize a compromised photomultiplier tube via error codes and based on various combinations of positive and indeterminate results. If a positive result with the negative control is obtained or false positive suspected, users are to contact Technical Service and cease using the instrument. The company is developing software solution aimed at identifying photomultiplier tubes that are at risk for failure and is seeking a hardware fix to this issue.
Quantity
452 units (398 US, 54 OUS)
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OR, PA, PR, RI, SD, TN, TX, UT, VA, VT, WI, WV and the countries of Australia, Chile, Italy, Switzerland.
Lot, serial or model codes
UDI: 00840733102318 All units in the field, all serial numbers.
510(k) numbers
["K170558"]
PMA numbers
not on file
Product code
OOI
Firm FEI
1524213
Firm city
Cincinnati
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register