FDA Medical Device Recalls (openFDA)
Animas Corporation: Animas(R) 2020, Animas(R) IR 1250 (1200 plus), Animas(R) IR 1200 Insulin Infusion Pumps Indicated for continuous subcutaneous infusion of insulin.
- Recall number
- Z-0836-2013
- Recalling firm
- Animas Corporation
- Product
- Animas(R) 2020, Animas(R) IR 1250 (1200 plus), Animas(R) IR 1200 Insulin Infusion Pumps Indicated for continuous subcutaneous infusion of insulin.
- Status
- Terminated
- Initiated
- 2012-12-17
- Posted
- 2013-02-17
- Terminated
- 2013-12-24
- Reason
- Due to an internal calendar date limit (December 31, 2015) these pumps will cease to operate on January 1, 2016, and will result in a call service alarm that is displayed on the pump display.
- Root cause
- Device Design
- Action
- The firm initiated their recall of this product on December 17, 2012 by sending a letter to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the Animas Customer Support Department at (866) 796-6373 for questions or concerns.
- Quantity
- 37,998
- Distribution
- Worldwide Distribution-USA (nationwide) including Washington, DC, Puerto Rico, US Virgin Island and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY, and the countries of Austria, Australia, Bermuda, Canada, New Zealand, Czech Republic, Italy, France, Finland, Hungary, Israel, United Kingdom, Spain, Sweden, Norway, Denmark, Germany, Mexico and Ireland.
- Lot, serial or model codes
- All Animas 2020, IR1250 and IR1200 Pump Models - Part Numbers 101370-00, 101371-00, 101372-00, 101374-00, 101375-00, 100380-00, 100381-00, 100382-00, 100384-00, 100385-00, 100450-00, 100210-00, 100211-00, 100212-00, 100250-00, 100251-00, 100252-00, 100200-00, 100201-00, 100202-00, 100170-00, 100171-00 and 100172-00.
- 510(k) numbers
- ["K032257","K042873"]
- PMA numbers
- not on file
- Product code
- LZG
- Firm FEI
- 3002920072
- Firm city
- West Chester
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28