Edwards Lifesciences, LLC: FORE-SIGHT ELITE Tissue Oximeter module, REF: HEMFSM10
- Recall number
- Z-0835-2022
- Recalling firm
- Edwards Lifesciences, LLC
- Product
- FORE-SIGHT ELITE Tissue Oximeter module, REF: HEMFSM10
- Status
- Open, Classified
- Initiated
- 2022-03-01
- Posted
- not on file
- Terminated
- not on file
- Reason
- The StO2 values may be inaccurately low when using either the FORE-SIGHT ELITE Tissue Oximeter Module or the FORE-SIGHT ELITE Absolute Tissue Oximeter Monitor with the Fore-Sight Elite large sensor in certain somatic locations (arms and legs). While the StO2 absolute values are impacted, the directional trend remains accurate, but may have a larger magnitude change. Low StO2 values may lead to unintended or inappropriate treatment
- Root cause
- Software in the Use Environment
- Action
- Urgent Medical Device Correction communication dated March 01, 2022 was sent via FedEx 2nd Day Advice on action to be taken by user: Customers are to refrain from using the large sensors on certain somatic locations (arms and legs). Cerebral and flank/abdomen locations for the large sensors are not impacted. Instructions for customers: " Review the customer letter for advice on how to use impacted product. " Please follow the instructions included in the enclosed acknowledgement form to complete the acknowledgement process. " Do not return any product. " Distribute this notice within your organization or to any organization where the potentially impacted product has been transferred. " Verify your inventory. " E-mail the completed form to Edwards Technical Support at tech_support@edwards.com, within 15 days from receipt of this notification. Advice on action to be taken by Distributor: Please review this letter and complete the acknowledgement form. Return the acknowledgement form to tech_support@edwards.com within 15 days of receipt of this notification. Please forward this customer communication to any of your customers who have purchased the impacted Edwards product. If you still have product within your control, please do not distribute impacted product to any customers. You can call Edwards Tech Support for instructions for handling product still within your control. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product from you and Edwards will communicate directly with your customers to facilitate the correction and acknowledgement process. Please forward your customer list to tech_support@edwards.com. If you have any questions, please contact Edwards Tech Support at 1 (800)-822-9837 If you have questions, contact Edwards Tech Support at 1 (800)-822-9837 from 5:00AM--4:30PM PST.
- Quantity
- 8389 units
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Brunei, Bulgaria, Canada, CANARY ISLANDS, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Northern Ireland, Norway, Oman, Panama, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates (UAE), United Kingdom, Vietnam.
- Lot, serial or model codes
- FORE-SIGHT ELITE Tissue Oximeter module, REF: HEMFSM10, UDI: 00690103208573; All serial numbers.
- 510(k) numbers
- ["K180003","K201446"]
- PMA numbers
- not on file
- Product code
- MUD
- Firm FEI
- 2015691
- Firm city
- Irvine
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28