FDA Medical Device Recalls (openFDA)
Ventana Medical Systems Inc: VENTANA (a member of the Roche Group) DISCOVERY ULTRA; REF/MU 750-601; Research Use Only Product Usage: Research Use Only-The DISCOVERY ULTRA Advanced Staining System is intended to automatically stain histological or cytological specimens on microscope slides with specific inununohistochemistry or in situ hybridization reagents.
- Recall number
- Z-0834-2013
- Recalling firm
- Ventana Medical Systems Inc
- Product
- VENTANA (a member of the Roche Group) DISCOVERY ULTRA; REF/MU 750-601; Research Use Only Product Usage: Research Use Only-The DISCOVERY ULTRA Advanced Staining System is intended to automatically stain histological or cytological specimens on microscope slides with specific inununohistochemistry or in situ hybridization reagents.
- Status
- Terminated
- Initiated
- 2012-09-28
- Posted
- 2013-08-06
- Terminated
- 2016-02-24
- Reason
- Potential for a waste fluid overflow condition to accur while running decontamination cycles on the BenchMark ULTRA automated staining platform or the DISCOVERY ULTRA automated staining platform.The waste fluid drainage system may become restricted by build-up of debris in the tubing and in-line filter, thereby inhibiting optimal flow rates. This type of overflow event can lead to migration of flu
- Root cause
- Device Design
- Action
- The firm, Roche sent a "Technical Bulletin" and an letter dated October 3, 2012 to its customers. The letter and bulletin described the product, problem and actions to be taken. The customers were informed that Ventana is aggressively working on solutions to this issue. Until mitigations to the overflow condition are available for implementation; inform all personnel in your organization that work with the BenchMark ULTRA or DISCOVERY ULTRA instrument about the potential for fluid overflow to occur during decontamination cycles; follow the instructions provided in the letter to avoid the documented failure mode and follow the instructions provided in the technical bulletin. Contact your local support center with any questions or concerns or call 520-877-7191.
- Quantity
- 54 instruments
- Distribution
- Worldwide distribution: USA (nationwide) and country of: Germany.
- Lot, serial or model codes
- Cat#750-601; PN 05987750001; Serial #'s: 310646, 311573, 310477, 311725, 311042, 311623, 312032, 312159, 310615, 310837, 311862, 311043 311572, 311766, 310526, 311011, 311185, 311723, 312082, 310635, 311574, 312084, 310636, 310835, 311179, 310493, 311183, 311769, 311930, 311962, 311963, 310437, 311965, 310997, 311726, 311620, 311512, 311526, 312083, 310468, 310759 (All serial numbers in the market).
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- KPA
- Firm FEI
- 1000125183
- Firm city
- Oro Valley
- Firm state
- AZ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28