Ventana Medical Systems Inc: VENTANA (a member of the Roche Group) DISCOVERY ULTRA; REF/MU 750-601; Research Use Only Product Usage: Research Use Only-The DISCOVERY ULTRA Advanced Staining System is intended to automatically stain histological or cytological specimens on microscope slides with specific inununohistochemistry or in situ hybridization reagents.

Recall number
Z-0834-2013
Recalling firm
Ventana Medical Systems Inc
Product
VENTANA (a member of the Roche Group) DISCOVERY ULTRA; REF/MU 750-601; Research Use Only Product Usage: Research Use Only-The DISCOVERY ULTRA Advanced Staining System is intended to automatically stain histological or cytological specimens on microscope slides with specific inununohistochemistry or in situ hybridization reagents.
Status
Terminated
Initiated
2012-09-28
Posted
2013-08-06
Terminated
2016-02-24
Reason
Potential for a waste fluid overflow condition to accur while running decontamination cycles on the BenchMark ULTRA automated staining platform or the DISCOVERY ULTRA automated staining platform.The waste fluid drainage system may become restricted by build-up of debris in the tubing and in-line filter, thereby inhibiting optimal flow rates. This type of overflow event can lead to migration of flu
Root cause
Device Design
Action
The firm, Roche sent a "Technical Bulletin" and an letter dated October 3, 2012 to its customers. The letter and bulletin described the product, problem and actions to be taken. The customers were informed that Ventana is aggressively working on solutions to this issue. Until mitigations to the overflow condition are available for implementation; inform all personnel in your organization that work with the BenchMark ULTRA or DISCOVERY ULTRA instrument about the potential for fluid overflow to occur during decontamination cycles; follow the instructions provided in the letter to avoid the documented failure mode and follow the instructions provided in the technical bulletin. Contact your local support center with any questions or concerns or call 520-877-7191.
Quantity
54 instruments
Distribution
Worldwide distribution: USA (nationwide) and country of: Germany.
Lot, serial or model codes
Cat#750-601; PN 05987750001; Serial #'s: 310646, 311573, 310477, 311725, 311042, 311623, 312032, 312159, 310615, 310837, 311862, 311043 311572, 311766, 310526, 311011, 311185, 311723, 312082, 310635, 311574, 312084, 310636, 310835, 311179, 310493, 311183, 311769, 311930, 311962, 311963, 310437, 311965, 310997, 311726, 311620, 311512, 311526, 312083, 310468, 310759 (All serial numbers in the market).
510(k) numbers
not on file
PMA numbers
not on file
Product code
KPA
Firm FEI
1000125183
Firm city
Oro Valley
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register