Skip to the content

FDA Medical Device Recalls (openFDA)

Hu-Friedy Mfg Co, Inc.: HENRY SCHEIN(R) 100-8008 Explorer-Double End #5. Barcode *+H658100800808*. Made in China. For professional use only. Instruments must be cleaned and sterilized prior to use. Distributed by: HENRY SCHEIN, INC. 135 Duryea Road, Melville, NY 11747 USA. This product is a dental explorer made of stainless steel packaged individually within a sleeve and containing a cardboard insert (labeling). 510 (k) exempt, Class I device, Device Listing # B057701.

Recall number
Z-0834-04
Recalling firm
Hu-Friedy Mfg Co, Inc.
Product
HENRY SCHEIN(R) 100-8008 Explorer-Double End #5. Barcode *+H658100800808*. Made in China. For professional use only. Instruments must be cleaned and sterilized prior to use. Distributed by: HENRY SCHEIN, INC. 135 Duryea Road, Melville, NY 11747 USA. This product is a dental explorer made of stainless steel packaged individually within a sleeve and containing a cardboard insert (labeling). 510 (k) exempt, Class I device, Device Listing # B057701.
Status
Terminated
Initiated
2004-03-29
Posted
2004-07-20
Terminated
2012-08-02
Reason
Reports of breakage on some dental explorers were received by the own label distributor and forwarded to Hu-Friedy Mfg. Co. (importer of record & Spec. Setter). Investigation conducted by Hu-Friedy Mfg. Co. revealed that the tips of the instruments may have been improperly heat treated and have the potential for unexpected tip fracture.
Root cause
Other
Action
On 3/29/04, Hu-Friedy Mfg. Co. notified Henry Schein, Inc. of the recall by telephone. This was followed up with a letter on 4/2/04. Henry Schein initiated a sub-recall to its customers by sending recall letters and response forms (dated 3/31/04) by First Class Mail on 4/2/04 & 4/5/04. The mailing will be followed up with fax, e-mail or phone calls.
Quantity
not on file
Distribution
The products were sold to the sole direct account: the Own Label Distributor, Henry Schein, Inc. in Melville, NY. Henry Schein further distributed the explorers to end users nationwide and internationally. Henry Schein Divisions located in Canada, Spain, France, Germany, United Kingdom, Holland, and New Zealand who further distributed the explorers to their customers in Canada and abroad.
Lot, serial or model codes
Date Codes/Lot Numbers: 0902, 1202, 0103 (0902 represents September 2002, 1202 represents December 2002, & 0103 represents January 2003); Product Code/Mfr. Part Code 100-8008.
510(k) numbers
not on file
PMA numbers
not on file
Product code
EKB
Firm FEI
1416605
Firm city
Chicago
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register