Ge Healthcare It: Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data. DICOM devices (e..g. modalities, workstations) Communicate with the archive using the DICOM protocol (published by ACR-NEMA). The clinical use is limited to the interaction that other systems have with the Centricity Enterprise archive. The intended users for Centricity Enterprise Archive are service personnel and system administrators and non clinical users. As of V4.0 XDS enabled systems can communicate with the archive using the XDS and XDS-1 profiles (published by IHE).

Recall number
Z-0833-2015
Recalling firm
Ge Healthcare It
Product
Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data. DICOM devices (e..g. modalities, workstations) Communicate with the archive using the DICOM protocol (published by ACR-NEMA). The clinical use is limited to the interaction that other systems have with the Centricity Enterprise archive. The intended users for Centricity Enterprise Archive are service personnel and system administrators and non clinical users. As of V4.0 XDS enabled systems can communicate with the archive using the XDS and XDS-1 profiles (published by IHE).
Status
Terminated
Initiated
2013-04-19
Posted
2014-12-19
Terminated
2016-06-10
Reason
GE Healthcare has become aware of a potential data loss issue associated with empty path names resulting from HL-7 updates when the cache disk is almost full.
Root cause
Software design
Action
General Electric sent an Urgent Medical Device Correction letter dated April 19, 2013, to all affected customers. The letter instructs customers to ensure that Enterprise Archive partitions have enough free space available, in particular the cache disk should have at least 10% free space available. This could be facilitated by setting appropriate disk limits and watermarks on the individual libraries in Enterprise Archive. In case the disks are not being purged to the configured watermarks and the cache disk is almost full, customers are requested to contact their local GE service person. A permanent product correction will be provided to all customers under FMI85209R. Customers with questions were instructed to contact their local Sales/Service Representative. Also the Remote Online Center (ROC) Customer Care can be reached at 1-800-437-1171.
Quantity
115 units
Distribution
Nationwide Distribution including AZ, OH, WA.
Lot, serial or model codes
Centricity Enterprise Archive versions: 4.0.x
510(k) numbers
not on file
PMA numbers
not on file
Product code
LMB
Firm FEI
3004526608
Firm city
Barrington
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register