FDA Medical Device Recalls (openFDA)
Ge Healthcare It: Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data. DICOM devices (e..g. modalities, workstations) Communicate with the archive using the DICOM protocol (published by ACR-NEMA). The clinical use is limited to the interaction that other systems have with the Centricity Enterprise archive. The intended users for Centricity Enterprise Archive are service personnel and system administrators and non clinical users. As of V4.0 XDS enabled systems can communicate with the archive using the XDS and XDS-1 profiles (published by IHE).
- Recall number
- Z-0833-2015
- Recalling firm
- Ge Healthcare It
- Product
- Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data. DICOM devices (e..g. modalities, workstations) Communicate with the archive using the DICOM protocol (published by ACR-NEMA). The clinical use is limited to the interaction that other systems have with the Centricity Enterprise archive. The intended users for Centricity Enterprise Archive are service personnel and system administrators and non clinical users. As of V4.0 XDS enabled systems can communicate with the archive using the XDS and XDS-1 profiles (published by IHE).
- Status
- Terminated
- Initiated
- 2013-04-19
- Posted
- 2014-12-19
- Terminated
- 2016-06-10
- Reason
- GE Healthcare has become aware of a potential data loss issue associated with empty path names resulting from HL-7 updates when the cache disk is almost full.
- Root cause
- Software design
- Action
- General Electric sent an Urgent Medical Device Correction letter dated April 19, 2013, to all affected customers. The letter instructs customers to ensure that Enterprise Archive partitions have enough free space available, in particular the cache disk should have at least 10% free space available. This could be facilitated by setting appropriate disk limits and watermarks on the individual libraries in Enterprise Archive. In case the disks are not being purged to the configured watermarks and the cache disk is almost full, customers are requested to contact their local GE service person. A permanent product correction will be provided to all customers under FMI85209R. Customers with questions were instructed to contact their local Sales/Service Representative. Also the Remote Online Center (ROC) Customer Care can be reached at 1-800-437-1171.
- Quantity
- 115 units
- Distribution
- Nationwide Distribution including AZ, OH, WA.
- Lot, serial or model codes
- Centricity Enterprise Archive versions: 4.0.x
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LMB
- Firm FEI
- 3004526608
- Firm city
- Barrington
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28