Cook Biotech, Inc.: Cook Medical Biodesign Otologic Butterfly Graft, REF ENT-OTO-BFLY-0.4-0.6, REF G60285, Sterile.

Recall number
Z-0832-2025
Recalling firm
Cook Biotech, Inc.
Product
Cook Medical Biodesign Otologic Butterfly Graft, REF ENT-OTO-BFLY-0.4-0.6, REF G60285, Sterile.
Status
Open, Classified
Initiated
2024-11-22
Posted
2025-01-10
Terminated
not on file
Reason
Incorrect version of IFU was used rather than updated version; and incorrect expiration date was used on two of the products.
Root cause
Under Investigation by firm
Action
The recalling firm issued two different emails on 11/22/2024, one informing the direct account that the devices either had an incorrect IFU revision (Biodesign Otologic Butterfly Graft, ENT-OTO-BFLY-0.4-0.6) or had an incorrect expiration date (Biodesign Hiatal Hernia Graft, C-PHR, and Biodesign Dural Graft, C-DUR); and the other informing a different direct account that the product had an incorrect IFU revision for two lots (for Biodesign Otologic Butterfly Graft, ENT-OTO-BFLY-0.4-0.6). The direct accounts informed the recalling firm, and the recalling firm issued "URGENT: MEDICAL DEVICE RECALL" letters to its customers dated 12/10/2024. The letters instructed the customers to examine inventory and quarantine affected product, return the products to Cook Biotech Inc with a copy of a completed "Acknowledge and Receipt Form even if you do not have affected product(s) on hand; share notice and immediately report adverse events to Cook Biotech by phone 765-807-7888 or by email to: biotech.feedback@rtix.com. If you have any questions or concerns, please contact Cook Biotech at 765-807-7888.
Quantity
49 devices
Distribution
US (domestic) Distribution to states of IN and MI; and OUS (foreign) to Canada and South Korea.
Lot, serial or model codes
Lot #LB1576761, exp. 1/21/2025; and Lot #LB1577308, exp. 1/21/2025; Box UDI-DI 10827002602851, Pouch UDI-DI 00827002602854.
510(k) numbers
["K232646"]
PMA numbers
not on file
Product code
KHJ
Firm FEI
3002340648
Firm city
W Lafayette
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register