FDA Medical Device Recalls (openFDA)

InfuTronix LLC: Nimbus PainPro Ambulatory Infusion Pumps with software version v4, Model Number IT1072, sold as kit IT1075-DUKE-HSP, IT1076-DUKE-HSP and IT1077-DUKE-HSP

Recall number
Z-0832-2022
Recalling firm
InfuTronix LLC
Product
Nimbus PainPro Ambulatory Infusion Pumps with software version v4, Model Number IT1072, sold as kit IT1075-DUKE-HSP, IT1076-DUKE-HSP and IT1077-DUKE-HSP
Status
Open, Classified
Initiated
2021-10-19
Posted
not on file
Terminated
not on file
Reason
An unreleased Version 4 Protocol Library was installed on Nimbus PainPro (H) infusion pumps in error that may impact patient safety if used on patients.
Root cause
Employee error
Action
An IMPORTANT PRODUCT INFORMATION letter dated 10/19/21 was sent to customers. InfuTronix LLC has recently become aware of an unreleased Version 4 Protocol Library for Duke University Hospital was installed on your Nimbus PainPro (H) infusion pumps in error that may impact patient safety if used on patients. Please ensure that all your potential customers and potential users in your facility are made aware of this notification and the recommended actions. Product Issue Version 4 of the Duke Protocol Library was installed on your infusion pumps in error. Affected Product Details Nimbus PainPro (H) Pump, Model numbers: IT1072, sold as kits IT1075-DUKE-HSP Total number of infusion pumps affected: 3 Lot numbers involved: 2009231440 Instructions 1. Place all units you have in-house on product-hold and stop further distribution 2. Inform your customer to discontinue use of the affected infusion pumps. Product Correction InfuTronix will be contacting you to resolve this issue as quickly as possible. Contact Information Your Account Manager, Vann Griffin, email: vann.griffin@nimbuspainpro.com, phone: (919) 606-8339 Or John LaFratta, Vice President of Sales, email: john.lafratta@ nimbuspainpro.com , phone: 315-729-5564. Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact us immediately.
Quantity
74 units
Distribution
US Nationwide distribution in the states of NC and OH.
Lot, serial or model codes
(UDI): (01)00817170020086; Lot: 2009231440; Serial Numbers: 807483, 807484, 807485, 807486, 807487, 807489, 807490, 807491, 807492, 807493, 807494, 807495, 807496, 807497, 807498, 807499, 807500, 807502, 807651, 807652, 807653, 807507, 807508, 807509, 807510, 807511, 807512, 807513, 807514, 807515, 807516, 807517, 807518, 807519, 807520, 807521, 807522, 807523, 807524, 807525, 807526, 807527, 807528, 807529, 807530, 807531, 807532, 807533, 807534, 807535, 807536, 807537, 807538, 807539, 807540, 807541, 807542, 807543, 807544, 807545, 807546, 807547, 807548, 807549, 807550, 807551, 807552, 807553, 807554, 807654, 807655, 807656, 807657, 807658
510(k) numbers
["K153193"]
PMA numbers
not on file
Product code
FRN
Firm FEI
3011581906
Firm city
Natick
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register