FDA Medical Device Recalls (openFDA)
Integra LifeSciences Corp.: Integra Licox Brain Tissue Oxygen Monitoring Integra Licox Recon Tissue Oxygen Monitoring 1) LICOX Complete Brain Probe Kit (comprised of IP1, CC1.P1) 2) LICOX Complete Brain Probe Kit (comprised of IP2, CC1.P1) 3) LICOX Complete Brain Tunnelling Probe Kit (comprised of VK5.2, CC1.P1 4) LICOX Complete Brain Probe Kit (comprised of IM1, CC1.SB) Sterile Rx Only Manufacturer CMS -Gesse;schaft for medizinische Sondentechnik mbH Dorfstrasse 2 24247
- Recall number
- Z-0831-2015
- Recalling firm
- Integra LifeSciences Corp.
- Product
- Integra Licox Brain Tissue Oxygen Monitoring Integra Licox Recon Tissue Oxygen Monitoring 1) LICOX Complete Brain Probe Kit (comprised of IP1, CC1.P1) 2) LICOX Complete Brain Probe Kit (comprised of IP2, CC1.P1) 3) LICOX Complete Brain Tunnelling Probe Kit (comprised of VK5.2, CC1.P1 4) LICOX Complete Brain Probe Kit (comprised of IM1, CC1.SB) Sterile Rx Only Manufacturer CMS -Gesse;schaft for medizinische Sondentechnik mbH Dorfstrasse 2 24247
- Status
- Terminated
- Initiated
- 2014-11-07
- Posted
- 2014-12-18
- Terminated
- 2015-08-11
- Reason
- Due to a complaint, it was determined that multiple lots of Licox catheters manufactured since October 29, 2013 contained a smart card which was programmed with an inaccurate lkN2 value.
- Root cause
- Device Design
- Action
- Integra sent Urgent- Voluntary Medical Device Recall Letters/Recall Acknowledgement and Return Form dated 11/7/2014 to customers and distributors via traceable courier service.
- Quantity
- 2,506 units
- Distribution
- Nationwide, Foreign
- Lot, serial or model codes
- CC1 Licox¿ PO2 Microprobe, 200 mm 021213, 040414, 070314, 130114, 240214, 270114, 280314, 300614 CC12 Licox¿ PO2 Microprobe, 300 mm 020614, 040814, 050514, 070414, 070714, 220414 CC1.P1 Licox¿ PMO Combined Probe Containing Oxygen and Temperature 050614, 060214, 060314, 100414, 100714, 130214, 130314, 170114, 190614, 210814, 260614, 310714 IM3ST Licox¿ Complete Brain IMC-Probe Kit: 020614, 030314, 110714, 130114, 140314, 160614, 180714, 200114, 280314 IP1.P Licox¿ Single Lumen Bolt CMPS and Temperature Bolt Kit 030714, 060214, 060314, 070814, 080514, 100714, 120614, 130314, 190614, 310714 IP2.P Licox¿ Double Lumen Bolt CMPS, Temperature and ICP Monitoring Kit 030714, 050614, 060214, 060314, 070814, 080514, 090114, 100714, 130314, 140814, 200214, 200414, 260614, 290814 IT2 Licox¿ Tunneling CMPS and Temperature Monitoring Kit 090114, 100414, 190614, 200214, 260614, 310714
- 510(k) numbers
- ["K002765","K040235","K053244"]
- PMA numbers
- not on file
- Product code
- GWM
- Firm FEI
- 3003418325
- Firm city
- Plainsboro
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28