Steris Corporation: Medisafe Distal Duck Kit and Duck Bag, Cleaner, ultrasonic, medical instrument - Product Usage: intended to keep instruments moist during transportation between point of use and cleaning as an initial step of reprocessing.
- Recall number
- Z-0830-2021
- Recalling firm
- Steris Corporation
- Product
- Medisafe Distal Duck Kit and Duck Bag, Cleaner, ultrasonic, medical instrument - Product Usage: intended to keep instruments moist during transportation between point of use and cleaning as an initial step of reprocessing.
- Status
- Terminated
- Initiated
- 2020-11-19
- Posted
- not on file
- Terminated
- 2023-10-25
- Reason
- STERIS has identified that certain lots of diluted 4-Zyme may contain bacteria, specifically Pseudomonas fluorescens. The presence of this bacteria can cause the color of the detergent to darken over time. There is an improbable risk to users of the product from exposure to this bacteria, and no risk to patients.
- Root cause
- Device Design
- Action
- On 11/23/2020, STERIS issued Urgent Device Correction notices to customers via courier service. Instructing customers to inspect and destroy any remaining affected products in their inventory. STERIS will provide replacement product at no cost to Customers who have product remaining in stock.
- Quantity
- 23,445 units, Domestic 5,727 units, Foreign 17,718 units
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of CA, DE, CO, VA, FL, WA, TX, OH, PA, MA, IA, CT and the countries of United Kingdom, Saudi Arabia, Portugal, Denmark, Ireland, United Arab Emirates, South Korea, Australia, Greece, Ashmore & Cartier Islands, Belgium, Canada, China, Germany, Japan, Norway, Poland, Qatar, Romania, Sweden, Thailand.
- Lot, serial or model codes
- Model# M20350 Lot# 1811273,1811277,1906484,1908593,1909635,1910651, 1911658,2001730,2004770 Model# M20358 Lot# 1903408,1906508,1907556,1909644,2001718,2004772 Model# M20359 Lot# 1811278,1906499,1910649,1911680,1911691,1912708, 2001719,2004768 Model# M20400 Lot# 1901340,1901349,1902374,1904448,1904464,1905505, 1906529,1907571,1908595,1909624,1910653,1912700, 2001717,2004766,2004774,2006797
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FLG
- Firm FEI
- 1527821
- Firm city
- Mentor
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28