Stryker Endoscopy: Stryker Infravision Esophageal Kit, Model Numbers 220180548, 220180540, 220180560, manufactured by Stryker Endoscopy San Jose.

Recall number
Z-0830-2011
Recalling firm
Stryker Endoscopy
Product
Stryker Infravision Esophageal Kit, Model Numbers 220180548, 220180540, 220180560, manufactured by Stryker Endoscopy San Jose.
Status
Terminated
Initiated
2009-08-20
Posted
2010-12-28
Terminated
2011-01-03
Reason
The silicone sheath component of the product detaches from the catheter during removal of the e-kit from the body.
Root cause
Device Design
Action
Stryker issued Urgent Product Recall letters dated August 20, 2009 to all its direct consignees, informing them of the affected products and providing instructions on the recall. Customers were requested to return the affected product to Styrker.
Quantity
2985 units
Distribution
Worldwide Distribution: USA in the states of: Alabama, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, Missouri, North Carolina, New Hampshire, Nebraska, New Jersey, New Mexico, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Virginia, Vermont, Washington, Wisconsin, West Virginia, and Wyoming, and the countries/region of Australia, Canada, Scandinavia, Italy, and Switzerland.
Lot, serial or model codes
5687,5766,5815,5822,5833,5872, 5924,60973888,60987814, 61004943, STROll409-l, STROl1409-2, STRO 11609-1, STROl1609-2, STR02l009-1, STR02l009-8, STR022709-l, STR022709-2, STR03ll09-1, STR03ll09-3, STR040l09-l, STR041709-2, STR041709-4, STR0430009-2, STRlO0208-l, STR102208-1, STR102408-l, STRI02708-7, STR103l08-1, STRlll308-l, STR12l808-l, STR12l808-2, STROl1209-l
510(k) numbers
["K983220"]
PMA numbers
not on file
Product code
FAT
Firm FEI
2936485
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register