Skip to the content

FDA Medical Device Recalls (openFDA)

Nova Biomedical Corporation: Stat Profile Critical Care Xpress CCX+ (R) Anaylzer Catalog NUmber: 37555 R= Refurbished

Recall number
Z-0829-04
Recalling firm
Nova Biomedical Corporation
Product
Stat Profile Critical Care Xpress CCX+ (R) Anaylzer Catalog NUmber: 37555 R= Refurbished
Status
Terminated
Initiated
2004-03-30
Posted
2004-07-20
Terminated
2004-06-17
Reason
Incorrectly assembled pumps may result in inconsistent compression of the tubing causing erratic results or calibration problems
Root cause
Other
Action
Nova Biomedical notified users by telephone on 3/31/04 and conducted follow-up by letter either by email or fax. On-site visit will be scheduled to replace the pump assembly.
Quantity
7units
Distribution
CA, IL, LA, MA ,NJ, NY, OH, PA, TX International: Italy ,England, Japan,Turkey., U.A.E., Greece, Thailand, Switzerland, Finland, Saudi Arabia, Italy, Mexico, Trinidad, Ireland, France, Cyprus, Czech Republic, ,Germany
Lot, serial or model codes
Domestic S/N: Y02902141 Y0290211Z Y02103141 Y02303101 Y02503281  International S/N: Y02603111 Y02603091
510(k) numbers
not on file
PMA numbers
not on file
Product code
CHL
Firm FEI
1219029
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register