NxStage Medical, Inc.: NXStage PureFlow SL SAK-303 - 2 dialysate sacks (SAK) with concentrate for use with PureFlow SL, Rx The NxStage PureFLow SL is an optional accessory to the NxStage System One that prepares dialysate for use during hemodialysis, as prescribed by the physician.

Recall number
Z-0826-2012
Recalling firm
NxStage Medical, Inc.
Product
NXStage PureFlow SL SAK-303 - 2 dialysate sacks (SAK) with concentrate for use with PureFlow SL, Rx The NxStage PureFLow SL is an optional accessory to the NxStage System One that prepares dialysate for use during hemodialysis, as prescribed by the physician.
Status
Terminated
Initiated
2011-12-14
Posted
2012-01-20
Terminated
2012-04-17
Reason
Report of high conductivity failure alarms with the PureFlow SL due to excess concentrate by weight in the product (SAK) dialysate Preparation System
Root cause
Employee error
Action
On December 13, 2011, NxStage sent replacement product for overnight delivery and initiated phone contact with all consignees of the product . In addition, NxStage sent an "IMPORTANT MEDICAL DEVICE RECALL" notice dated December 14, 2011 to its consignees/customers. The notice describes the product, problem and actions to be taken. The customers were instructed to use SAKs from a different lot or premixed dialysate bags (RFPs), until their replacement SAKs arrived; set aside all SAKs from listed lot number; record their inventory; dispose of the SAKs following their normal practices, and complete and return the DEVICE PRODUCT RECALL REPLY FORM even if they do not have any of the affected product on hand via postage-prepaid envelop provided. If you have any questions about this notice, contact NxStage Customer Support at 866-Nxstage ( 697-8243).
Quantity
1,192 devices
Distribution
Nationwide distribution.
Lot, serial or model codes
Lot 1117911, Exp. 2013-09
510(k) numbers
["K080919","K111174"]
PMA numbers
not on file
Product code
FKR
Firm FEI
3003464075
Firm city
Lawrence
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register