Skip to the content

FDA Medical Device Recalls (openFDA)

Cincinnati Sub-Zero Products LLC, a Gentherm Company: Hemotherm CE Dual Reservoir Cooler/Heater, Model No. 400CE Product Usage: The HEMOTHERM Model 400CE Dual Reservoir Cooler/Heater is used to lower, maintain, or raise the temperature of the water flowing through a Blood Oxygenator / Heat Exchanger that is used to cool or warm blood during cardiopulmonary bypass procedures lasting six hours or less. The Hemotherm Dual Reservoir Cooler/Heater may also be used with a hyper / hypothermia blanket under the patient to provide warming through conductive heat transfer.

Recall number
Z-0825-2019
Recalling firm
Cincinnati Sub-Zero Products LLC, a Gentherm Company
Product
Hemotherm CE Dual Reservoir Cooler/Heater, Model No. 400CE Product Usage: The HEMOTHERM Model 400CE Dual Reservoir Cooler/Heater is used to lower, maintain, or raise the temperature of the water flowing through a Blood Oxygenator / Heat Exchanger that is used to cool or warm blood during cardiopulmonary bypass procedures lasting six hours or less. The Hemotherm Dual Reservoir Cooler/Heater may also be used with a hyper / hypothermia blanket under the patient to provide warming through conductive heat transfer.
Status
Terminated
Initiated
2018-12-19
Posted
not on file
Terminated
2020-04-24
Reason
Evidence supports that when customers performed a fuse replacement as part of a March 2017 field action, the fuse clips became stretched and were no longer mechanically tight with the fuse, creating a host spot that led to discoloration on the board due to overheating and shut down of the device.
Root cause
Device Design
Action
Initial notification letters were sent to customers on 12/19/18 to inform them of the upcoming field action. An additional letter sent to customers on 1/18/19 providing instructions on the correction. The letter identified the affected product, problem and actions to be taken.
Quantity
396
Distribution
Worldwide Distribution - US Nationwide in the states of: AL, AZ, CA, CO, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, VA, and WI. The products were distributed to the following foreign countries: Columbia and Taiwan.
Lot, serial or model codes
Serial numbers 151-10283CE through 171-10721CE
510(k) numbers
K122813
PMA numbers
not on file
Product code
DWC
Firm FEI
1516825
Firm city
Cincinnati
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source