Skip to the content

FDA Medical Device Recalls (openFDA)

Tosoh Bioscience Inc: ST AIA-PACK PROG II, Progesterone II Assay, Part Number: 025239

Recall number
Z-0825-2018
Recalling firm
Tosoh Bioscience Inc
Product
ST AIA-PACK PROG II, Progesterone II Assay, Part Number: 025239
Status
Terminated
Initiated
2018-02-13
Posted
not on file
Terminated
2020-10-29
Reason
Falsely elevated progesterone assay values will occur when using certain progesterone assay part numbers due to cross reactivity of DHEA-S, a metabolite of the DHEA (dehydroepiandrosterone) supplement.
Root cause
Device Design
Action
The assays identified in the above table must not be used to report results for patients who are taking DHEA supplements. " For patients that are taking DHEA supplements, an alternative testing method which is not expected to show cross reactivity to DHEA-S should be used such as Liquid Chromatography-Mass Spectroscopy (LCMS). " Please review the information in this Urgent Medical Device Recall with your Medical Director and/or Lab Director, as soon as possible. " Continue to use the Progesterone Assays for patients who are not receiving DHEA supplements or therapy. " Complete and return the attached Urgent Medical Device Recall Acknowledgement Form within 15-days. " Maintain this notification with your laboratory records and forward this information to others who may have received this product. " If you have obtained unexpected test results or received any complaints of illness or adverse events associated with the use of the products identified in the table above, contact Tosoh Technical Support 24 hours a day, seven days a week at 1(800) 248-6764 or email Bernadette.OConnell@tosoh.com. " Should you have any questions regarding this medical device recall, please feel free to contact Bernadette OConnell at 1(800) 248-6764, or by email at Bernadette.OConnell@tosoh.com. We will be available to answer any questions Monday - Friday, from 8:00 AM to 5:00 PM (PST).
Quantity
5579
Distribution
Distributed in 39 states: AK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI.
Lot, serial or model codes
All lots
510(k) numbers
not on file
PMA numbers
not on file
Product code
JLS
Firm FEI
3005529799
Firm city
Grove City
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register