FDA Medical Device Recalls (openFDA)
Trumpf Medical Systems, Inc.: Camera System, Television, Surgical without audio Intended for transmission and documentation of surgical interventions for the purpose of archiving and teaching.
- Recall number
- Z-0824-2018
- Recalling firm
- Trumpf Medical Systems, Inc.
- Product
- Camera System, Television, Surgical without audio Intended for transmission and documentation of surgical interventions for the purpose of archiving and teaching.
- Status
- Terminated
- Initiated
- 2017-12-15
- Posted
- not on file
- Terminated
- 2020-04-23
- Reason
- The radiated emission from the TruVidia" Wireless Receiver is above the normative accepted limits, in which the radiated emission may interact with other devices in the operating room and cause them not to function properly.
- Root cause
- Other
- Action
- Trumpf Medical Systems sent an Urgent Medical Device Correction letter dated December 15, 2017. Instructions include to complete and return the Confirmation of Receipt, arrange for the Trumpf Medical Engineer or an authorized Service Engineer to upgrade the affected device, and ensure that all users of the device are aware of the recall. For further questions, please call (843) 329-0543
- Quantity
- 212 devices
- Distribution
- Worldwide Distribution - US Distribution to the states of : GA, IL, IN, PA, SC, TX, UT, VA. and to the countries of : Austria, Australia, Bangladesh, Belgium, France, Switzerland, Czech Republic, Germany, Denmark, Spain, United Arab Emirates, United Kingdom, Ireland, Italy, Kuwait, Lithuania, Poland, Portugal, Romania, Sweden, Slovenia, Thailand, Turkey.
- Lot, serial or model codes
- Catalog Number 1940747, All serial numbers: 102584474, 102584475, 102584477, 102584478, 102805330, 102813656, 102813657, 102813658, 102813659, 102813660, 102813661, 102813662, 102860647, 102860652, 102863706, 102863707, 102867783, 102867784, 102925734, 102925735, 102925736, 102925737, 102925738, 102925739, 102925740, 102925741, 102925742, 102925743, 102928703, 102932185, 102932186, 102932187, 102932188, 102932189, 102932203, 102932204, 102932205, 102932206, 102932207, 102932208, 102932209, 102932210, 102932211, 102932212, 102932217, 102932218, 102932219, 102932220, 102932221, 102932223, 102934525, 102934526, 102934527, 102934528, 102934529, 102936117, 102939484, 102940382, 102940383, 102940384, 102940386, 102940387, 102940388, 102940389, 102940390, 102940391, 102940408, 102940409, 102940410, 102961390, 102961391, 102961392, 102961393, 102961394, 102961395, 102961396, 102961397, 102961398, 102961399, 102961400, 102961401, 102975336, 102986494, 102986495, 102988480, 102989096, 102989607, 102989608, 102989615, 102990285, 102990286, 102994502, 102994503, 102994504, 102994505, 102994506, 102994525, 102994526, 102994529, 102994530, 102996306, 103005879, 103005880, 103005889, 103005890, 103005891, 103005892, 103005893, 103006636, 103010023, 103010759, 103010760, 103010761, 103010762, 103010763, 103012786, 103012787, 103013644, 103017040, 103017041, 103019900, 103020713, 103022348, 103022349, 103022350, 103022351, 103022352, 103022794, 103024049, 103024050, 103030284, 103031623, 103034106, 103034109, 103038842, 103038843, 103038844, 103038845, 103038846, 103038847, 103038848, 103038849, 103038850, 103038853, 103038854, 103038855, 103038856, 103038857, 103038858, 103038859, 103038860, 103038861, 103038862, 103038863, 103038864, 103038865, 103038866, 103039468, 103042319, 103042320, 103042321, 103042322, 103042925, 103042926, 103046119, 103046120, 103046121, 103046122, 103046123, 103046124, 103046125, 103046126, 103046127, 103046128, 103046129, 103046130, 103048248, 103065061, 103070795, 103070796, 103070797, 103070798, 103079001, 103079053, 103087365, 103087366, 103087367, 103087368, 103087369, 103087370, 103087371, 103090135, 103090136, 103090137, 103091507, 103091508, 103091509, 103091510, 103091511, 103095664, 103101676, 103101826, 103101864, 103101886, 103103841, 103103842, 103103843, 103103844, 103103845, 103104556, 103113059, 103117961, 103118986, 103118987, 103118988, 103118989, 103118990, 103118991, 103118992, 103118993, 103118994, 103118995, 103123759, 103133224, 103134656, 103138822, 103143214, 103149993, 103164298, 103168053, 103168341, 103168431, 103176559, 103176571, 103176574, 103191728, 103191729, 103194419, 103214005, 103214022.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FWB
- Firm FEI
- 3003184737
- Firm city
- Charleston
- Firm state
- SC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28