Siemens Healthcare Diagnostics, Inc.: Siemens CentraLink" Data Management System V14.0.4, V14.0.5 or V14.0.8 The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location.
- Recall number
- Z-0823-2014
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Product
- Siemens CentraLink" Data Management System V14.0.4, V14.0.5 or V14.0.8 The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location.
- Status
- Terminated
- Initiated
- 2013-10-30
- Posted
- 2014-01-24
- Terminated
- 2014-12-15
- Reason
- Under extremely rare circumstances, a patient result that had been previously disabled may be released without proper review. (Disabled Test in CentraLink System or Aptio System May Become Enabled in CentraLink System.)
- Root cause
- Software design
- Action
- The firm, Siemens, sent an "Urgent Medical Device Correction" letter dated October 23, 2013 to its customers. The firm also sent a follow-up letter dated October 30, 2013. The letter describes the product, problem and actions to be taken. The customers were instructed Until a service visit is scheduled, do the following: Instruct the laboratory staff to recognize this sequence of events and Ensure that whenever the operator disables a method on an Aptio system or a CentraLinksystem and replenishes reagent for that method, the operator then confirms that the method remains disabled in the CentraLink system or disables it again. The operator must allow up to five minutes to verify that the test remains disabled. In addition, the customers were instructed tocomplete and return the FIELD CORRECTION EFFECTIVENESS CHECK response form via fax to Siemens Healthcare Diagnostics Customer Care Center at 800-441-9250, and to retain this letter with your laboratory records, and forward this letter to those who may have received this product. After a service visit is performed and an interim solution is implemented, the laboratory staff will be trained on this update to understand the minor but necessary changes in the workflow. If you have any questions, please contact your Siemens Customer Care Center at 800-441-9250 or your local Siemens representative.
- Quantity
- 40
- Distribution
- US Distribution in the states of: MN, NJ, and NV.
- Lot, serial or model codes
- CentraLink" Data Management System V14.0.4, V14.0.5 or V14.0.8 Catalog Numbers: 10810210, 10811596, 10814296, 10814298, 10814877, 10815474.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JQP
- Firm FEI
- 2517506
- Firm city
- Newark
- Firm state
- DE
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28