Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: Device Brand Name: Sirius Polish Cemented Stem 34B Device Classification Name: prosthesis, hip, semi-constrained, metal/polymer, cemented Device Common Name: Sirius Femoral Stem The Biomet's Sirius Femoral Hip Stem is a double-tapered, highly polished cemented hip stem, designed for either primary or revision surgery. A distal centralizer is provided for optimal placement within the femoral canal. The system is used with Type 1 taper modular heads and compatible acetabular shells/liners and screws to provide a total hip replacement system.

Recall number
Z-0822-2015
Recalling firm
Biomet, Inc.
Product
Device Brand Name: Sirius Polish Cemented Stem 34B Device Classification Name: prosthesis, hip, semi-constrained, metal/polymer, cemented Device Common Name: Sirius Femoral Stem The Biomet's Sirius Femoral Hip Stem is a double-tapered, highly polished cemented hip stem, designed for either primary or revision surgery. A distal centralizer is provided for optimal placement within the femoral canal. The system is used with Type 1 taper modular heads and compatible acetabular shells/liners and screws to provide a total hip replacement system.
Status
Terminated
Initiated
2014-12-02
Posted
2014-12-17
Terminated
2015-10-27
Reason
After receiving a complaint, Biomet Orthopedic Inc., conducted an investigation which determined that the following parts were mixed up prior to packaging and shipment: Sirius Polish Cemented Stem 34B Catalog Number: 51-199333, Lot Number Identification: 024320 and Sirius Polish Cemented Stem 38C, Catalog Number: 51-199335, Lot Number Identification: 141680, 180920.
Root cause
Mixed-up of materials/components
Action
Sent notices to customers dated 12/2/2014 notifying them of the recall: 5 were implanted, 3 were shipped to subsidiary for testing, and 2 are in Biomet control. Public Contact for the recall is: Audrey Daenzer, Field Action Specialist, Regulatory Compliance audrey.daenzer@biomet.com, can be contacted from Monday through Friday 8 AM to 5 PM, 574-372-1570.
Quantity
9
Distribution
Foreign Distribution in China and Australia only.
Lot, serial or model codes
Sirius Polish Cemented Stem 348 Catalog Number: 51-199333 Lot Number Identification: 024320
510(k) numbers
["K130610"]
PMA numbers
not on file
Product code
JDI
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register