Skip to the content

FDA Medical Device Recalls (openFDA)

Veridex, LLC: CellSearch Circulating Tumor Cell Kit (Epithelial) IVD REF 7900001 Veridex LLC, a Johnson & Johnson Company.

Recall number
Z-0819-2010
Recalling firm
Veridex, LLC
Product
CellSearch Circulating Tumor Cell Kit (Epithelial) IVD REF 7900001 Veridex LLC, a Johnson & Johnson Company.
Status
Terminated
Initiated
2009-08-12
Posted
2010-02-23
Terminated
2012-07-18
Reason
Veridex has received 4 complaints regarding reagent bottles having been incorrectly placed or duplicate bottles places within the reagent tray of CellSearch Circulating Tumor Cell Kit. Delayed detection of disease relapse or progression may occur due to false low CTC results.
Root cause
Other
Action
Urgent Product Correction Notification letters were sent to all customers on August 12, 2009 by Federal Express. Customers were advised to inspect the kits to assure the proper reagents are in the proper position. If not, do not use the kits and contact the company at 877-837-4339 (chose option 1).
Quantity
1641 in US, 380 internationally.
Distribution
Product was distributed worldwide, US (one government account: National Cancer Institute, Frederick, MD). Also 8 international accounts in Belgium, France, Germany, Italy, Spain, United Kingdom, Netherlands, Norway and Israel.
Lot, serial or model codes
Lot number 0019, Exp 1/8/2010; 0019B, Exp 1/8/2010; 0039, Exp 3/16/2010; 0059, Exp 3/6/2010, 0079, Exp 2/20/2010; 0099, Exp 6/24/2010.
510(k) numbers
["K073338"]
PMA numbers
not on file
Product code
NQI
Firm FEI
3004619490
Firm city
Raritan
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register