Smith & Nephew, Inc., Endoscopy Div.: High Flow Irrigation Extender; Part Number 72201021. Arthroscopic accessories are indicated for use in access of the joint capsule.

Recall number
Z-0818-2016
Recalling firm
Smith & Nephew, Inc., Endoscopy Div.
Product
High Flow Irrigation Extender; Part Number 72201021. Arthroscopic accessories are indicated for use in access of the joint capsule.
Status
Terminated
Initiated
2013-07-30
Posted
not on file
Terminated
2021-02-24
Reason
During arthroscopic surgery, when the scope is moved inside the cannula, the body of the irrigation extender separates, potentially obstructing visualization and causing leakage of the irrigation fluid. The device is removed and a back-up device is used to complete the surgery.
Root cause
Under Investigation by firm
Action
Urgent - Product Recall 1st Notification - Urgent R-2013-13 letters, dated July 30, 2013 were sent overnight via Federal Express to customers. The letters identified the affected product, as well as the potential risk involved. Actions to be taken were provided for both hospital and sales representatives. Hospital representatives were asked to inspect their inventory, locate all devices from the list provided, and quarantine them immediately. The Inventory Return Certification Form should be completed and returned, regardless of whether any affected product is on hand or not. Contact the Smith & Nephew Returns Group at 800-343-5717 option 3) or send an e-mail to ProductRecovery@smith-nephew.com. They will provide instructions for returning affected product and receiving a replacement.
Quantity
203 units
Distribution
Worldwide Distribution -- US, including the states of AL, AR, CA, FL, GA, ID, IL, KS, MA, MI, MO, NC, TN, TX, and AK; and, country of Japan.
Lot, serial or model codes
50209581 50220887 50231401 50216929 50226220 50231932 50218285 50229266 50235063 50240019 50245683 50251109 50254562 50256942 50263784 50279382 50291593 50302984 50303418 50327069 50330377 50351345 50361381 50404796 50408768 50419699 50433964 50250837 50256854 50269580 5027766 50304642 50342002 50401593 50417254
510(k) numbers
not on file
PMA numbers
not on file
Product code
NBH
Firm FEI
1219602
Firm city
Mansfield
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register