Skip to the content

FDA Medical Device Recalls (openFDA)

DePuy Spine, Inc.: Depuy Spine Mountaineer OCT Spinal System Screw Caddy Set, , Non Sterile Product Code: 2883-90-400 The MOUNTAINEER Screw caddy is intended to act as a holder in an implant set for implantable screws prior to use in surgery. Further, the caddy provides a means to organize the screws in accordance with the length of the screws.

Recall number
Z-0818-2011
Recalling firm
DePuy Spine, Inc.
Product
Depuy Spine Mountaineer OCT Spinal System Screw Caddy Set, , Non Sterile Product Code: 2883-90-400 The MOUNTAINEER Screw caddy is intended to act as a holder in an implant set for implantable screws prior to use in surgery. Further, the caddy provides a means to organize the screws in accordance with the length of the screws.
Status
Terminated
Initiated
2010-10-02
Posted
2010-12-27
Terminated
2011-04-15
Reason
The scale which serves as a secondary means of identifying screw length is incorrectly etched on the screw caddy
Root cause
Labeling design
Action
Depuy Spine issued an "URGENT-VOLUNTARY PRODUCT RECALL NOTIFICATION" via e-mail on October 2, 2010 to customers. The notification describes the product, problem and action to be taken by customers. The customers were instructed to please distribute this information to appropriate personnel at their facility and return any caddies that they may have to their local distributor for DePuy Spine products. Depuy will return devices to customer promptly (within two business days), after reworking the devices by removing the scale altogether. Once the product is fully corrected, Depuy will provide the customers with a new and updated unit with the corrected scale markings. If you should have any questions or concerns regarding this action, please contact DePuy Spine Customers Service at 800-227-6633.
Quantity
121 units
Distribution
Nationwide distribution: USA including states of: AR, AZ, CA, FL, HI, IA, IL, KY, LA, MA, MI, MN, MO, NC, NY OH, PA, TN, TX, VA, and WA.
Lot, serial or model codes
Not lot coded
510(k) numbers
not on file
PMA numbers
not on file
Product code
LRP
Firm FEI
1526439
Firm city
Raynham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register