American Contract Systems, Inc.: Custom procedural convenience kits and trays, cardiac, labeled as: a) HEART CATHETERIZATION TRAY, kit number JRHC45G; b) TAVR PACK, kit number LLTV10; c) CATH PED LF SJH, kit number SJCA21F; d) CATH HEART ADL LF SJH, kit number SJCH22F; e) INT RAD PORTS-LF - 206060, kit number UDIR58AL
- Recall number
- Z-0817-2024
- Recalling firm
- American Contract Systems, Inc.
- Product
- Custom procedural convenience kits and trays, cardiac, labeled as: a) HEART CATHETERIZATION TRAY, kit number JRHC45G; b) TAVR PACK, kit number LLTV10; c) CATH PED LF SJH, kit number SJCA21F; d) CATH HEART ADL LF SJH, kit number SJCH22F; e) INT RAD PORTS-LF - 206060, kit number UDIR58AL
- Status
- Open, Classified
- Initiated
- 2023-11-29
- Posted
- 2024-01-26
- Terminated
- not on file
- Reason
- During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
- Root cause
- Process control
- Action
- Owens & Minor issued an URGENT: MEDICAL DEVICE CORRECTION notice to it consignees on 11/29/2023 via email. Then notice explained the problem, risk to health, and requested the following actions be taken: If you received any of the batches/lots identified, please segregate and quarantine all affected product. Add warning labels to all kits/cases impacted: A labeling template is enclosed for printing warning labels which should be affixed to all affected product in inventory. This label should be affixed to each affected kit and case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. " Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. " Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified of this field correction and maintain records of effectiveness. Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com.
- Quantity
- 4394 units
- Distribution
- US Nationwide distribution.
- Lot, serial or model codes
- a) JRHC45G, UDI/DI 00191072164240, Lot Numbers: 912231, exp. 3/29/2024; 920231, exp. 3/21/2024; 952231, exp. 2/17/2024; 963231, exp. 2/6/2024; 976231, exp. 6/2/2023; 654221, exp. 12/12/2023; 714221, exp. 10/13/2023; 717221, exp. 6/7/2023; 766221, exp. 8/22/2023; 814221, exp. 7/5/2023; b) LLTV10, UDI/DI 00191072183255, Lot Numbers: 644221, exp. 12/22/2023; 973231, exp. 1/27/2024; 952231, exp. 2/17/2024; 924231, exp. 3/17/2024; c) SJCA21F, UDI/DI None, Lot Numbers: 666221, exp. 11/30/2023; 701221, exp. 10/26/2023; 747221, exp. 9/10/2023; 779221, exp. 8/9/2023; 807221, exp. 7/12/2023; 846221, exp. 6/3/2023; 919231, exp. 3/22/2024; 954231, exp. 2/15/2024; 988231, exp. 1/12/2024; d) SJCH22F, UDI/DI None, Lot Numbers: 670221, exp. 11/26/2023; 708221, exp. 10/19/2023; 747221, exp. 9/10/2023; 795221, exp. 7/24/2023; 834221, exp. 6/15/2023; 873221, exp. 5/7/2023; 937231, exp. 3/4/2024; 991231, exp. 1/9/2024; e) UDIR58AL, UDI/DI None, Lot Numbers: 682221, exp. 11/14/2023; 638221, exp. 7/31/2023; 963231, exp. 2/6/2024; 925231, exp. 3/16/2024; 883231, exp. 10/30/2023
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OES
- Firm FEI
- 3003309947
- Firm city
- Temple Terrace
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28