Skip to the content

FDA Medical Device Recalls (openFDA)

Remel, Inc: Oxoid dry Spot Staphytect Plus, Catalog #DR0100M, 120 tests/kit. Oxoid Ltd., Wade road, Basingstoke, Hants, UK. Latex slide agglutination test for the differentiation of staphylococcus aureus by detection of clumping factor, Protein A and certain polysaccarides found in methicillin-resistant S. aureus (MRSA) from those staphylococci that do not possess these properties.

Recall number
Z-0817-2010
Recalling firm
Remel, Inc
Product
Oxoid dry Spot Staphytect Plus, Catalog #DR0100M, 120 tests/kit. Oxoid Ltd., Wade road, Basingstoke, Hants, UK. Latex slide agglutination test for the differentiation of staphylococcus aureus by detection of clumping factor, Protein A and certain polysaccarides found in methicillin-resistant S. aureus (MRSA) from those staphylococci that do not possess these properties.
Status
Terminated
Initiated
2009-07-20
Posted
2010-02-23
Terminated
2012-04-25
Reason
The products may give false positive results with some non-Staphylococcus aureus strains
Root cause
Process control
Action
Thermo Fischer Scientific, Remel Inc. issued letters dated July 20, 2009 via regular mail explaining the reason for recall, risks involved, and requested they discard all remaining product. A Product Inventory Checklist was enclosed to report the amount of product their customer discarded. A second letter dated October 19, 2009 was issued via certified mail. For further information, contact Thermo Fischer Scientific, Remel Inc. at 1-800-255-6730, Extension 3304185.
Quantity
27 kits
Distribution
Nationwide Distribution -- MN, CT, FL, GA, CA, NY, MI, MO, PA, and IL.
Lot, serial or model codes
Lot Number: J055610, Exp June 2011
510(k) numbers
not on file
PMA numbers
not on file
Product code
JWX
Firm FEI
1924669
Firm city
Lenexa
Firm state
KS
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register