Hitachi Medical Systems America Inc: Ultrasound Transducer, Model Number: UST-5550-R, is used in conjunction with a diagnostic ultrasound system evaluation during robotic and non-robotic intra-operative and laparoscopic procedures.

Recall number
Z-0816-2018
Recalling firm
Hitachi Medical Systems America Inc
Product
Ultrasound Transducer, Model Number: UST-5550-R, is used in conjunction with a diagnostic ultrasound system evaluation during robotic and non-robotic intra-operative and laparoscopic procedures.
Status
Terminated
Initiated
2017-10-20
Posted
not on file
Terminated
2020-04-16
Reason
The ultrasound probe may not have adequate protection against electrical shock hazards.
Root cause
Mixed-up of materials/components
Action
Hitachi must recall the lot of UST-5550-R Transducers from S/N 1 to S/N 155. If you have a serial number in this range, please contact Hitachi Service at (800) 800-4925 to schedule the recall. Hitachi Sales will contact you regarding the options for a replacement transducer.
Quantity
55
Distribution
Distributed in 21 states: AR, AZ, CA, CO, FL, GA, IA, IL, IN, MA, MO, NC, NY, OK, OR, PA, SC, TX, VA, WA, WV.
Lot, serial or model codes
17 21 32 37 38 39 41 43 45 46 48 57 58 59 61 63 64 69 73 77 78 81 82 86 87 92 96 97 99 107 108 109 114 115 116 118 120 121 122 126 129 130 132 135 139 140 141 143 144 147 148 151 155
510(k) numbers
["K152126"]
PMA numbers
not on file
Product code
ITX
Firm FEI
1528028
Firm city
Twinsburg
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register