Skip to the content

FDA Medical Device Recalls (openFDA)

Alpha Omega Services Inc: VariSource AL 13199024, Implant Tube, Implant Tube, Intralumenal(HDR), 10CM Drawdown 130CM Body, Sealed End, Cable Insert & Gold Seed, Disposable

Recall number
Z-0816-06
Recalling firm
Alpha Omega Services Inc
Product
VariSource AL 13199024, Implant Tube, Implant Tube, Intralumenal(HDR), 10CM Drawdown 130CM Body, Sealed End, Cable Insert & Gold Seed, Disposable
Status
Terminated
Initiated
2006-02-21
Posted
2006-04-27
Terminated
2006-08-30
Reason
There is a possibility that the diameter of the catheter does not meet specifications. As a result it may not be possible to correctly connect the catheter to a VariSource Quick Connect. This would prevent any treatment. Furthermore, since the problem would not be apparent until after the catheter has been implanted an unnecessary surgical implantation procedure would have taken place.
Root cause
Other
Action
Recall Notification Letter: began February 22, 2006. Depth/Scope: 100% of the customer''s who purchased the affected products will be identified and contacted with a faxed and mailed Medical Device Recall notice and followed up with a phone call by the AOS Sales & Marketing personnel. All customers will be requested return all affected products, both used and unused.
Quantity
24
Distribution
Nationwide and Argentina, Brasil, Canada, Israel, Korea, , Taiwan, United Kingdom & Venezuela
Lot, serial or model codes
All Lots
510(k) numbers
not on file
PMA numbers
not on file
Product code
JAQ
Firm FEI
2022694
Firm city
Bellflower
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register