Skip to the content

FDA Medical Device Recalls (openFDA)

St. Jude Medical, Atrial Fibrillation Division, Inc.: TactiFlex Ablation Sensor Enabled Ablation Catheter, Models A-TFSE-FJ, 7.5 F shaft and an 8F distal section, Sterile.

Recall number
Z-0814-2024
Recalling firm
St. Jude Medical, Atrial Fibrillation Division, Inc.
Product
TactiFlex Ablation Sensor Enabled Ablation Catheter, Models A-TFSE-FJ, 7.5 F shaft and an 8F distal section, Sterile.
Status
Open, Classified
Initiated
2023-12-18
Posted
2024-01-25
Terminated
not on file
Reason
When the catheters are used with the EnSiteX EP System, the system does not correctly recognize the catheter as the appropriate product model and instead reads and displays it as a different model. This results in some features specific to the catheter do not appear on the display and the catheter image may also appear inverted.
Root cause
Under Investigation by firm
Action
The recalling firm issued letters dated 12/18/2023 via hand-delivery and FedEx beginning 12/18/2023. The letter explained the issue, the scope of the problem, the impact and associated risks, and next steps to help reduce risk. These steps were: (1) Do not use any remaining inventory from the affected lots listed in Appendix A; (2) Complete and return the accompanying Acknowledgment Form; (3) Return all remaining unused devices from the affected lots; and (4) Forward the letter to anyone within their organization who may need to be notified. Replacements of returned affected devices with a similar curve are available and the customer's Abbott representative can assist in returning these devices and obtaining replacements.
Quantity
483 units
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, DC, DE, FL, GA, IA, IL, KS, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, and WV. The country of Canada.
Lot, serial or model codes
Lot numbers 10073391, 10075618, 10080798, 10086492, 10087953, 10089624, 10092157, 10093066, and 10095363, Exp. 11/30/2024, GTIN 05415067034618.
510(k) numbers
not on file
PMA numbers
["P130026"]
Product code
OAE
Firm FEI
1000139754
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register