Skip to the content

FDA Medical Device Recalls (openFDA)

Cytocell Ltd.: TCL1 Breakapart Probe, Fluorescence in situ hybridisation (FISH) is a technique allows the visualisation of DNA sequences upon chromosome Catalog Number: LPH046-A

Recall number
Z-0814-2022
Recalling firm
Cytocell Ltd.
Product
TCL1 Breakapart Probe, Fluorescence in situ hybridisation (FISH) is a technique allows the visualisation of DNA sequences upon chromosome Catalog Number: LPH046-A
Status
Terminated
Initiated
2022-02-04
Posted
not on file
Terminated
2024-11-06
Reason
May show unexpected locus specific signals in addition to those at 14q32.
Root cause
Under Investigation by firm
Action
Cytocell issued Urgent Medical Device Recall letter (VC/2022/005) to Distributor/End User via email on 2/4/22 . Letter states reason for recall, health risk and action to take: Immediately examine your inventory and quarantine all product subject to recall. Cytocell requests that you destroy the remaining inventory. We also suggest that laboratories undertake a review of the results obtained with the affected devices and check that signal patterns were not misinterpreted as a result of any additional locus specific signals at 9q34. Please complete and return the enclosed response form as soon as possible. If you have any questions, contact Cytocell on Tel: +44 (0)1223 294048, Fax +44(0)1223 294986 or vigilance@ogt.com.
Quantity
1 unit
Distribution
MN
Lot, serial or model codes
Lot 074612
510(k) numbers
not on file
PMA numbers
not on file
Product code
MVU
Firm FEI
3002646096
Firm city
Cambridge
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register