FDA Medical Device Recalls (openFDA)
Bard Access Systems: MRI Low Profile Single Lumen port with Open-Ended 6.6Fr Silicone Catheter. Product Code 0603880. Product Usage: The 6.6 Fr Open-Ended Single-Lumen venous catheters is packaged into various Bard Acess Systems finished goods as Attachable or Preconnected catheters in BardPort Implanted Port Kits These products are used for patient therapies requiring repeated access to the vascular system. They are indicated for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. Subject ports are not indicated to Power Injection. Port access is performed by percutaneous needle insertion using a non-coring needle. The system consists of two primary components: an injection port with a self-sealing silicone septum and a radiopaque silicone catheter.
- Recall number
- Z-0814-2014
- Recalling firm
- Bard Access Systems
- Product
- MRI Low Profile Single Lumen port with Open-Ended 6.6Fr Silicone Catheter. Product Code 0603880. Product Usage: The 6.6 Fr Open-Ended Single-Lumen venous catheters is packaged into various Bard Acess Systems finished goods as Attachable or Preconnected catheters in BardPort Implanted Port Kits These products are used for patient therapies requiring repeated access to the vascular system. They are indicated for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. Subject ports are not indicated to Power Injection. Port access is performed by percutaneous needle insertion using a non-coring needle. The system consists of two primary components: an injection port with a self-sealing silicone septum and a radiopaque silicone catheter.
- Status
- Terminated
- Initiated
- 2013-12-06
- Posted
- 2014-01-23
- Terminated
- 2014-08-05
- Reason
- Bard Access Systems is recalling various Low Profile Port with 6.6 Fr Single Lumen Broviac Catheter because a small number of Silicone catheters in the may be undersized to 5.3 Fr instead of 6.6 Fr as stated in the labeling for the device.
- Root cause
- Process control
- Action
- BARD sent an Urgent Product Recall letter dated October 2013 to affected customers via certified letter. The letter identified the affected product, problem and actions to be taken. Customers were asked to check all inventory locations, remove and return any unused affected prodcut to Bard Access Systems through return goods authorization number RAP80498 for a credit or replacement, and to complete the Reply Form and Inventory Reconcilation Form. For questions contact BAS Customer Service at 1-800-290-1689.
- Quantity
- 1358
- Distribution
- Worldwide Distribution - US nationwide, Belgium, and Canada.
- Lot, serial or model codes
- Lot REXC0931 and REXC0736
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FOZ
- Firm FEI
- 3006260740
- Firm city
- Salt Lake City
- Firm state
- UT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as