Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Inc. Cardiac Rhythm Disease Management: Custom Packs/Mexico, Medtronic Cardio-Kit, Sterilized Using Ethylene Oxide, Nonpyrogenic, Rx Only. Model # DLP6N48R3 & DLP6N48R4. Medtronic Intersept Custom Tubing Pack with Trillium Biosurface, Sterilized Using Ethylene Oxide, , Model # TL4F71R1. These cannulae are intended for collection of venous blood from the right side of the heart via the superior and inferior vena cava during cardiopulmonary bypass surgery up to six hours or less.

Recall number
Z-0813-2012
Recalling firm
Medtronic Inc. Cardiac Rhythm Disease Management
Product
Custom Packs/Mexico, Medtronic Cardio-Kit, Sterilized Using Ethylene Oxide, Nonpyrogenic, Rx Only. Model # DLP6N48R3 & DLP6N48R4. Medtronic Intersept Custom Tubing Pack with Trillium Biosurface, Sterilized Using Ethylene Oxide, , Model # TL4F71R1. These cannulae are intended for collection of venous blood from the right side of the heart via the superior and inferior vena cava during cardiopulmonary bypass surgery up to six hours or less.
Status
Terminated
Initiated
2011-12-14
Posted
2012-01-18
Terminated
2014-05-22
Reason
Instructions for Use for the DLP Single Stage Venous Cannulae are in the process of being updated. The current precaution statement referencing ventricular assist procedures will be identified as a contraindication in the updated Instructions for Use. DO NOT USE THIS PRODUCT FOR EXTENDED TERMS SUCH AS VENTRICULAR ASSIST PROCEDURES. Medtronic has received two reports of patient deaths due to p
Root cause
Labeling design
Action
Medtronic sent an "URGENT MEDICAL DEVICE INFORMATION" letter dated December 2011 to all affected customers. The letter identifies the product, problem, and the action to be taken by the customers. Contract Medtronic CardioVascular LifeLine Technical Support at 1-877-526-7890 for questions related to this issue.
Quantity
106
Distribution
Worldwide Distribution-USA (nationwide) including Puerto Rico and the Virgin Islands and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Barbados, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macau, Macedonia, Malaysia, Mauritius, Mexico, Montenegro, Morocco, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Suriname, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Viet Nam, and Yemen.
Lot, serial or model codes
Model # / Lot #    DLP6N48R3 DLP6N48R4 TL4F71R1 11241044 11465359 11799944  11755661  11713167  11713167  11669547  11616953  11568176  11509030  11478537  11432935  11414606  11282018  11252748  11239976  11209356  11130765  11110871  11101609  11051824  11015290  10916961  10860025  10816466  10693674  10656545  10588143  10502496  10454472  10426945  10310272  10255607   11669547  11616953  11568176  11509030  11478537  11432935  11414606  11282018  11252748  11239976  11209356  11130765  11110871  11101609  11051824  11015290  10916961  10860025  10816466  10693674  10656545  10588143  10502496  10454472  10426945  10310272  10255607
510(k) numbers
["K891687"]
PMA numbers
not on file
Product code
DTZ
Firm FEI
2182208
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register