FDA Medical Device Recalls (openFDA)
Spacelabs Healthcare, Llc: Blease Sirius Anaesthesia System is a anaesthesia Gas Machine. The system has the capability to mount two anaesthetic vapourisers on the back bar via a demountable system. The delivery of anaesthetic agent is achieved by the rotation of a control knob which actuates a feature designed "shot bold" mechanism. The purpose of which is to prevent two vapourisers being turned on simultaneously.. The product is labeled in part: "BleaseSirius Anesthesia System...SPACELABS Healthcare...Blease Medical Equipment Limited Beech House, Chiltern Court, Asheridge Road, Chesham, Bucks HP5 2PX, England...Spacelabs Healthcare Company Headquarters 5150 220th Ave. SE, Issaquah, WA 98029 PO Box 7018, Issaquah, WA 98027-7018".
- Recall number
- Z-0813-2011
- Recalling firm
- Spacelabs Healthcare, Llc
- Product
- Blease Sirius Anaesthesia System is a anaesthesia Gas Machine. The system has the capability to mount two anaesthetic vapourisers on the back bar via a demountable system. The delivery of anaesthetic agent is achieved by the rotation of a control knob which actuates a feature designed "shot bold" mechanism. The purpose of which is to prevent two vapourisers being turned on simultaneously.. The product is labeled in part: "BleaseSirius Anesthesia System...SPACELABS Healthcare...Blease Medical Equipment Limited Beech House, Chiltern Court, Asheridge Road, Chesham, Bucks HP5 2PX, England...Spacelabs Healthcare Company Headquarters 5150 220th Ave. SE, Issaquah, WA 98029 PO Box 7018, Issaquah, WA 98027-7018".
- Status
- Terminated
- Initiated
- 2009-06-25
- Posted
- 2010-12-27
- Terminated
- 2010-12-28
- Reason
- Users may experience difficulty turning the control knob to an active vapour delivery setting, or may not be able to turn on the Left Hand mounted vapouriser on the Blease Sirius Anaesthesia Systems.
- Root cause
- Other
- Action
- Spacelabs Healthcare issued an URGENT MEDICAL DEVICE CORRECTION letter dated June 25, 2009 to the two consignees in the US. Spacelabs Healthcare UK notified consignees outside the US. The consignees were advised to brief their staff about the issue and test the Anesthesia Systems in advance of a procedure. If there is an issue with turning on a vaporizer in the left hand position, simply move the vaporizer to the right hand position. The firm contacted their consignees for a schedule to replace the flowmeter mouldings. The US consignees can contact Spacelabs Healthcare Representatives at 425-657-7200 ext. 5089 about this recall. Foreign consignees can contact Spoacelabs at 44 1494 784422.
- Quantity
- 5 units to US consignees and 36 units Internationally.
- Distribution
- Worldwide Distribution: US in the states of LA and SD, and internationally by Spacelabs Healthcare, Ltd in the countries of China, Czech Republic, Hong Kong, India, Poland, Saudi Arabia, Singapore, and the United Kingdom.
- Lot, serial or model codes
- Serial numbers: SIRI-000845, SIRI-000862, SIRI-000869, SIRI-000885, and SIRI-000898.
- 510(k) numbers
- ["K051629"]
- PMA numbers
- not on file
- Product code
- BSZ
- Firm FEI
- 3010157426
- Firm city
- Issaquah
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28