Abbott Point of Care Inc.: Abbott i-STAT ACT Celite Cartridges. In vitro diagnostic test used to monitor moderate and high-level heparin therapy through analysis of arterial and venous whole blood samples.
- Recall number
- Z-0813-2010
- Recalling firm
- Abbott Point of Care Inc.
- Product
- Abbott i-STAT ACT Celite Cartridges. In vitro diagnostic test used to monitor moderate and high-level heparin therapy through analysis of arterial and venous whole blood samples.
- Status
- Terminated
- Initiated
- 2009-08-18
- Posted
- 2010-02-23
- Terminated
- 2010-05-18
- Reason
- Labeling discrepancy: The manual instruction state that the line should be flushed with 5 mL of saline and the first 5 mL of blood or six dead space volumes should be discarded. The Cartridge Test Information sheets that it should be 2 mL of blood - which is incorrect.
- Root cause
- Labeling False and Misleading
- Action
- Abbott Point of Care issued an "Important Product Information" letter beginning August 18, 2009. Consignees were instructed on how to reconcile all affected product. For further information, contact Abbott Point of Care Technical Support at 1-800-366-8020, Option 1.
- Quantity
- 696,675 Cartridges
- Distribution
- Worldwide distribution.
- Lot, serial or model codes
- Catalog number: 07G01-1
- 510(k) numbers
- ["K992571"]
- PMA numbers
- not on file
- Product code
- JBP
- Firm FEI
- 2245578
- Firm city
- Princeton
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28