Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF): vitatron (implantable pulse generator): (a) A-series, Model Numbers: A30A1, A60A1 (b) E-series, Model Numbers: E50A1, E60A1 (c) G-series, Model Numbers: G70A1, G70A2 (d) Q-series, Model Numbers: Q50A2, Q70A2, Q80A2

Recall number
Z-0812-2019
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Product
vitatron (implantable pulse generator): (a) A-series, Model Numbers: A30A1, A60A1 (b) E-series, Model Numbers: E50A1, E60A1 (c) G-series, Model Numbers: G70A1, G70A2 (d) Q-series, Model Numbers: Q50A2, Q70A2, Q80A2
Status
Open, Classified
Initiated
2019-01-17
Posted
2019-02-14
Terminated
not on file
Reason
A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that affects device functionality
Root cause
Component design/selection
Action
In the US, beginning 17-Jan-2019, Medtronic Field Representatives hand deliver an FCA Notification Letter to consignees including implanting and follow-up physicians. In consultation with an Independent Physician Quality Panel (IPQP), patient management recommendations were provided with the FCA Notification letter to ensure patient safety. Medtronic Field Representatives will hand deliver an additional Supplemental Letter with the Urgent Medical Device Recall letter to a subset of physicians with patients whose device has shown evidence of a pacing pause that is potentially related to this circuit error. Consignees will be asked to return all unused and unopened affected product to Medtronic.
Quantity
47086 units
Distribution
Wordlwide
Lot, serial or model codes
vitatron (implantable pulse generator):  (a) A-series, Model Number: A30A1, GTIN 00643169709331; A60A1, GTIN 00643169709294, All Serial Numbers (b) E-series, Model Number: E50A1, GTIN 00643169709126, 00643169709140; E60A1, GTIN 00643169709164, 00643169709188, All Serial Numbers (c) G-series, Model Numbers: G70A1, GTIN 00643169709249, 00643169709263; G70A2, GTIN 00643169883000, All Serial Numbers (d) Q-series, Model Number: Q50A2, GTIN 00643169883017; Q70A2, GTIN 00643169882980; Q80A2, GTIN 00643169882997, All Serial Numbers
510(k) numbers
not on file
PMA numbers
not on file
Product code
NVZ
Firm FEI
2182208
Firm city
Mounds View
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source