Skip to the content

FDA Medical Device Recalls (openFDA)

Medex Inc: MX6160R1, Custom CIPER (Comprehensive Intracranial Presssure Evaluation and Relief) System. The product is packaged in a tyvek pouch and is sold sterile. The product is labeled on the tyvek pouch with a stick-on label that reads in part ,''medex Dublin, OH 43016 USA***STERILE**MX6160R1 ** Custom CIPER**Lot #:**Latex Free***SINGLE USE''

Recall number
Z-0811-06
Recalling firm
Medex Inc
Product
MX6160R1, Custom CIPER (Comprehensive Intracranial Presssure Evaluation and Relief) System. The product is packaged in a tyvek pouch and is sold sterile. The product is labeled on the tyvek pouch with a stick-on label that reads in part ,''medex Dublin, OH 43016 USA***STERILE**MX6160R1 ** Custom CIPER**Lot #:**Latex Free***SINGLE USE''
Status
Terminated
Initiated
2005-11-18
Posted
2006-04-26
Terminated
2008-01-22
Reason
The product is labeled as 'Latex Free'; however, one of the components of the finished product may contain latex.
Root cause
Other
Action
The firm sent a recall letter dated 11/18/2005, via Certified mail to all of their customers. In the letter, the firm requests that the consignee telephone the firm and obtain a Return Authorization number in order to return the recalled product.
Quantity
2480 devices
Distribution
The recalled device was distributed to a single direct account customer located in Alabama and to
Lot, serial or model codes
All Lot codes listed as follows are affected: 30K200116, 31A160060, 31A250027, 31B280077, 31E160051,31H030002, 31I140023, 31J050052,31J250065, 32C070041, 32I050037, 32K250019, 33A300046, 33C170028, 33E290076, 33H220008, 3L170057, 34B16D038, 34B26D062, 34E05D055, 34F09D079, 34H05D023, 34H25D034, 34T17D091,34J14D021, 34L16D040
510(k) numbers
["K822864"]
PMA numbers
not on file
Product code
GWM
Firm FEI
1526863
Firm city
Dublin
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register