Skip to the content

FDA Medical Device Recalls (openFDA)

Heartport, Inc.: The Heartport EndoClamp aortic catheter is a 10.5 Fr. wirewound, three-lumen catheter with an elastomeric balloon near its tip. This device is used with a 200 cm j-hook guide wire accessory device.

Recall number
Z-0811-05
Recalling firm
Heartport, Inc.
Product
The Heartport EndoClamp aortic catheter is a 10.5 Fr. wirewound, three-lumen catheter with an elastomeric balloon near its tip. This device is used with a 200 cm j-hook guide wire accessory device.
Status
Terminated
Initiated
2005-04-12
Posted
2005-05-10
Terminated
2005-10-31
Reason
Guidewire is protruding through the film portion of the packages which compromises the sterility of the device.
Root cause
Other
Action
The recall was initiated on 4/8/2005. All accounts were phoned so as to cancel any scheduled surgical procedures that rely on these devices. Also the firm formally notified the hospitals in writing via letters to the Director of Material Managment, Director of OR, as well as the Director of the Cardiovascular Suite. Letters were sent via certified mail.
Quantity
1029 units
Distribution
The devices were distributed to 70 consignees in the US and 43 international customers. The devices are distributed directly to customers (hospitals, clinics) in the US, Belgium, Canada, Czech Republic, France, Germany, Great Britain, Hungary, Israel, Italy, Netherlands, and Spain. There are no government accounts.
Lot, serial or model codes
Product Code EC1001:  Lot Number, exp. date: MS020434, EXP APRIL-2005; MS0504003, EXP APRIL-2005; MS0504027, EXP MAY-2005; MS0604006, EXP MAY-2005; MS0604007, EXP MAY-2005; MS0704006, EXP JUNE-2005; MS0704015, EXP JUNE-2005; MS0804007, EXP AUGUST-2005; MS1004007, EXP OCTOBER-2005; MS1104021, EXP OCTOBER-2005; MS0105002, EXP JANUARY-2006; MS0105047, EXP FEBRUARY-2006; MS0205039, EXP FEBRUARY-2006; MS0205036, EXP MARCH-2006; MS0205041, EXP MARCH-2006; MS0205042, EXP MARCH-2006.
510(k) numbers
["K974175"]
PMA numbers
not on file
Product code
DWZ
Firm FEI
3000719637
Firm city
Somerville
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register