Skip to the content

FDA Medical Device Recalls (openFDA)

Spacelabs Healthcare, Inc.: Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data.

Recall number
Z-0810-2025
Recalling firm
Spacelabs Healthcare, Inc.
Product
Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data.
Status
Open, Classified
Initiated
2024-11-25
Posted
2025-01-07
Terminated
not on file
Reason
Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts and necessary clinical intervention could be delayed.
Root cause
Software change control
Action
On 11/25/2024, correction notices were emailed to customers who were asked to do the following: Firm will contact you at the earliest possible date to schedule a convenient time to update affected devices to software version 1.4.2 for XTR systems which addresses this defect. You can also contact GTS to have an FSE paged out for discussing upgrade schedule: 1-800-522-7025 and select 2 for Technical Support. Complete and submit the customer acknowledgement form: https://spacelabshealthcare.com/support/recall-reply-form/recall-reply-form-xtr-15-nov-2024/ For additional information or technical assistance, or to report an adverse event, contact: Technical Support: 1-800-522-7025 and select 2
Quantity
434
Distribution
US: MN, TX, TN, SC, OK, NM, GA, PA, OH, IA, AL, CO, AR, FL, MO, ME, LA, IL, WA, MA, WI, OR, MS, ID, NY, KS, PR. OUS: Canada, Taiwan, Czech Republic, France, Greece, Italy, Kuwait, Philippines, Poland.
Lot, serial or model codes
Software Versions: 1.4.0 and 1.4.1. UDI-DI: 10841522107177, 10841522100246
510(k) numbers
["K141156"]
PMA numbers
not on file
Product code
MHX
Firm FEI
3010157426
Firm city
Snoqualmie
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register