Medrad Mr Inc: Medrad Veris MR Monitor units The system is intended to monitor physiological parameters of patients within any healthcare environment. The user, responsible to interpret the monitored data made available, will be a professional healthcare provider. Physiological data, gas monitoring, system alarms, and patient analysis will be available to the care provider from the monitor.

Recall number
Z-0810-2014
Recalling firm
Medrad Mr Inc
Product
Medrad Veris MR Monitor units The system is intended to monitor physiological parameters of patients within any healthcare environment. The user, responsible to interpret the monitored data made available, will be a professional healthcare provider. Physiological data, gas monitoring, system alarms, and patient analysis will be available to the care provider from the monitor.
Status
Terminated
Initiated
2013-11-21
Posted
2014-01-23
Terminated
2015-03-30
Reason
The main board, P/N 301641, installed in some Medrad Veris MR Monitor units may be faulty and could lead to unexpected shutdown of the system while in use, resulting in the loss of vital signs information from the monitor.
Root cause
Device Design
Action
The firm, Bayer HealthCare, sent a "MEDICAL DEVICE FIELD CORRECTIVE ACTION" letter dated November 21, 2013 via certified mail to its users/customers. The letter described the product, problem and actions to be taken. The firm stated that, "You may choose to continue to use your Veris units but you MUST be vigilant about this potential issue. Should a shutdown occur, the system can be quickly restarted by pressing the POWER button. If this problem does occur, we ask that you report this immediately to Bayer Customer Support at 1-877-229-3767." The customers were instructed to verify the serial number on your Medrad Veris unit(s). If this problem does occur, report this immediately to Bayer Customer Support at 1-877-229-3767. The customers were also instructed to fill out and fax back the attached form. Bayer Service Representative will then be in touch to schedule an appointment to come and replace your main board, at no cost to you. Bayer also informed their customers that based on the maximum inventory that can be produced by the supplier, they anticipate that this Field Corrective Action will take up to six months. If you have questions, please contact our Customer Support team at 1-877-229-3767.
Quantity
922 units
Distribution
Worldwide Distribution: US (nationwide) and Internationally to: Canada.
Lot, serial or model codes
P/N 301641, Catalog #s 3011992, 3011992W, 3011993, 3011993W, 3011994, 3011994W, 3011996, 3011996W, 3014268, 3014268W, with multiple serial numbers.
510(k) numbers
["K042569"]
PMA numbers
not on file
Product code
MWI
Firm FEI
3006316319
Firm city
Warrendale
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register