Roche Diagnostics Corp.: Roche PTH (parathyroid hormone test system) for use in conjunction with Roche cobas e 411, cobas e 601, Elecsys 1010, Elecsys 2010 and Modular E 170 analyzers; catalog no. 11972103122.

Recall number
Z-0810-2007
Recalling firm
Roche Diagnostics Corp.
Product
Roche PTH (parathyroid hormone test system) for use in conjunction with Roche cobas e 411, cobas e 601, Elecsys 1010, Elecsys 2010 and Modular E 170 analyzers; catalog no. 11972103122.
Status
Terminated
Initiated
2007-04-19
Posted
2007-05-16
Terminated
2008-06-11
Reason
Negative bias caused by hemolysis interference at the low end of the reference range could cause erroneous diagnosis of hypoparathyroidism.
Root cause
Other
Action
The firm issued a letter dated 04/19/07 to all users informing them of the change in the hemoglobin interference claim.
Quantity
14,503
Distribution
Nationwide.
Lot, serial or model codes
Lots 17723501, exp. 1/31/08; 17723502, exp. 1/31/08; 17642601, exp. 11/30/07; 17642602, exp. 11/30/07; 17450901, exp. 5/31/07; 17450902, exp. 5/31/07 and 17450903, exp. 5/31/07.
510(k) numbers
["K992680"]
PMA numbers
not on file
Product code
CEW
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register