Skip to the content

FDA Medical Device Recalls (openFDA)

Philips North America: Achieva 3.0T, Model No. 781177

Recall number
Z-0809-2024
Recalling firm
Philips North America
Product
Achieva 3.0T, Model No. 781177
Status
Open, Classified
Initiated
2023-12-29
Posted
2024-01-25
Terminated
not on file
Reason
An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.
Root cause
Under Investigation by firm
Action
On December 29, 2023, the firm notified affected customers with URGENT Medical Device Correction letters. The firm stated there is no impact to patients or users. There is a potential risk of electrical shock for service personnel. Customers may continue to use MR systems in accordance with the intended use. Circulate this notice to all users of this device so that they are aware of the issue, and post the letter in the technical room until the electrical shock warning label is updated. A Philips representative will contact you to schedule a time for a Field Service Engineer to verify the electrical shock warning label and apply the correct label if applicable. If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).
Quantity
7 US; 16 ROW total
Distribution
Worldwide - US Nationwide distribution in the states of AR, CO, IL, MD, MS, NE, TX and the countries of Brazil, Canada, China, Germany, Greece, India, Mexico, and Saudi Arabia.
Lot, serial or model codes
Model No. 781177 UDI-DI N/A Serial Number 17043
510(k) numbers
["K193215"]
PMA numbers
not on file
Product code
LNI
Firm FEI
3006648320
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register